The application of extended modified Lambert Beer model for measurement of blood carboxyhemoglobin and oxyhemoglobin saturation 被引量:2
1
作者 Audrey Huong Xavier Ngu 《Journal of Innovative Optical Health Sciences》 SCIE EI CAS 2014年第3期108-114,共7页
This work presents the use of extended Modified Lambert Beer(MLB)model for accurate andcontinuous monitoring of percent blood carboxyhemoglobin(COHb)(SCO)and oxyhemoglobin(OxyHb)saturation(SO,)via a fitting procedure.... This work presents the use of extended Modified Lambert Beer(MLB)model for accurate andcontinuous monitoring of percent blood carboxyhemoglobin(COHb)(SCO)and oxyhemoglobin(OxyHb)saturation(SO,)via a fitting procedure.This quantification technique is based on theabsorption characteristics of hemoglobin derivatives in the wavelength range of 520-600 nm togive the best estimates of the required parameters.A comparison of the performance of the developed model and MILB law is made using attenuation data from Monte Carlo simulations for a two-layered skin model.The results revealed a lower mean absolute error of 0.4%in the valuesestimated by the developed model as compared to 10%that is given by the MILB law.This studyshowed that the discussed approach is able to provide consistent and accurate measurement ofblood SO,and SCO across diferent skin pigmen tations suggesting that it may potentially be usedas an alternative means for clinical diagnosis of carbon monoxide(CO)poisoning. 展开更多
关键词 Blood carboxyhemoglobin saturation carbon monoxide poisoning blood oxyhemoglobin saturation modified Lambert Beer law