期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
Long-term follow-up of right ventricular outflow tract septal pacing
1
作者 Xiaoqing Ren Shu Zhang Jielin Pu Fangzheng Wang 《Journal of Geriatric Cardiology》 SCIE CAS CSCD 2009年第2期71-74,共4页
Objective Right ventricular outflow tract septum has become widely used us an electrode placement site. However, data concerning lead performances and complications for lead repositioning with this technique were scan... Objective Right ventricular outflow tract septum has become widely used us an electrode placement site. However, data concerning lead performances and complications for lead repositioning with this technique were scant. The purpose of this study was to observe long- term lead performances and complications of right ventricular outflow tract septal pacing and provide evidences for choosing an optimal electrode implantation site. Methods Thirty-six patients with septal active electrode implantation and 39 with apical passive electrode implantation were enrolled in this study. Pacing threshold, R-wave sensing, lead impedance, pacing QRS width and pacing-related compli- cations for two groups at implantation and follow-up were compared. Results There were higher pacing threshold and shorter pacing QRS width at implantation in the septal group compared with the apical group. There were no differences between the septal and the apical groups in pacing threshold, R-wave sensitivity, lead impedance and pace-related complication during a follow-up. Conclusions Right ventricular outflow tract septum could be used as a first choice for implantation site because it had long-term stable lead performances and no serious complications compared with the traditional apical site. 展开更多
关键词 Cardiac pacemaker right ventricular outflow tract septum pacing threshold COMPLICATIONS
下载PDF
上一页 1 下一页 到第
使用帮助 返回顶部