The current parallel ankle rehabilitation robot(ARR)suffers from the problem of difficult real-time alignment of the human-robot joint center of rotation,which may lead to secondary injuries to the patient.This study ...The current parallel ankle rehabilitation robot(ARR)suffers from the problem of difficult real-time alignment of the human-robot joint center of rotation,which may lead to secondary injuries to the patient.This study investigates type synthesis of a parallel self-alignment ankle rehabilitation robot(PSAARR)based on the kinematic characteristics of ankle joint rotation center drift from the perspective of introducing"suitable passive degrees of freedom(DOF)"with a suitable number and form.First,the self-alignment principle of parallel ARR was proposed by deriving conditions for transforming a human-robot closed chain(HRCC)formed by an ARR and human body into a kinematic suitable constrained system and introducing conditions of"decoupled"and"less limb".Second,the relationship between the self-alignment principle and actuation wrenches(twists)of PSAARR was analyzed with the velocity Jacobian matrix as a"bridge".Subsequently,the type synthesis conditions of PSAARR were proposed.Third,a PSAARR synthesis method was proposed based on the screw theory and type of PSAARR synthesis conducted.Finally,an HRCC kinematic model was established to verify the self-alignment capability of the PSAARR.In this study,93 types of PSAARR limb structures were synthesized and the self-alignment capability of a human-robot joint axis was verified through kinematic analysis,which provides a theoretical basis for the design of such an ARR.展开更多
Aiming at the problem that the existing ankle rehabilitation robot is difficult to fully fit the complex motion of human ankle joint and has poor human-machine motion compatibility,an equivalent series mechanism model...Aiming at the problem that the existing ankle rehabilitation robot is difficult to fully fit the complex motion of human ankle joint and has poor human-machine motion compatibility,an equivalent series mechanism model that is highly matched with the actual bone structure of the human ankle joint is proposed and mapped into a parallel rehabilita-tion mechanism.The parallel rehabilitation mechanism has two virtual motion centers(VMCs),which can simulate the complex motion of the ankle joint,adapt to the individual differences of various patients,and can meet the reha-bilitation needs of both left and right feet of patients.Firstly,based on the motion properties and physiological structure of the human ankle joint,the mapping relationship between the rehabilitation mechanism and ankle joint is determined,and the series equivalent model of the ankle joint is established.According to the kinematic and con-straint properties of the ankle equivalent model,the configuration design of the parallel ankle rehabilitation robot is carried out.Secondly,according to the intersecting motion planes theory,the full-cycle mobility of the mechanism is proved,and the continuous axis of the mechanism is judged based on the constraint power and its derivative.Then,the kinematics of the parallel ankle rehabilitation robot is analyzed.Finally,based on the OpenSim biomechanical soft-ware,a human-machine coupling rehabilitation simulation model is established to evaluate the rehabilitation effect,which lays the foundation for the formulation of a rehabilitation strategy for the later prototype.展开更多
基金Supported by Key Scientific Research Platforms and Projects of Guangdong Regular Institutions of Higher Education of China(Grant No.2022KCXTD033)Guangdong Provincial Natural Science Foundation of China(Grant No.2023A1515012103)+1 种基金Guangdong Provincial Scientific Research Capacity Improvement Project of Key Developing Disciplines of China(Grant No.2021ZDJS084)National Natural Science Foundation of China(Grant No.52105009).
文摘The current parallel ankle rehabilitation robot(ARR)suffers from the problem of difficult real-time alignment of the human-robot joint center of rotation,which may lead to secondary injuries to the patient.This study investigates type synthesis of a parallel self-alignment ankle rehabilitation robot(PSAARR)based on the kinematic characteristics of ankle joint rotation center drift from the perspective of introducing"suitable passive degrees of freedom(DOF)"with a suitable number and form.First,the self-alignment principle of parallel ARR was proposed by deriving conditions for transforming a human-robot closed chain(HRCC)formed by an ARR and human body into a kinematic suitable constrained system and introducing conditions of"decoupled"and"less limb".Second,the relationship between the self-alignment principle and actuation wrenches(twists)of PSAARR was analyzed with the velocity Jacobian matrix as a"bridge".Subsequently,the type synthesis conditions of PSAARR were proposed.Third,a PSAARR synthesis method was proposed based on the screw theory and type of PSAARR synthesis conducted.Finally,an HRCC kinematic model was established to verify the self-alignment capability of the PSAARR.In this study,93 types of PSAARR limb structures were synthesized and the self-alignment capability of a human-robot joint axis was verified through kinematic analysis,which provides a theoretical basis for the design of such an ARR.
基金Supported by National Natural Science Foundation of China(Grant No.52075145)S&T Program of Hebei Province of China(Grant No.20281805Z)+1 种基金Hebei Provincial Natural Science Foundation of China(Grant No.E2022202130)Central Government Guides Basic Research Projects of Local Science and Technology Development Funds of China(Grant No.206Z1801G).
文摘Aiming at the problem that the existing ankle rehabilitation robot is difficult to fully fit the complex motion of human ankle joint and has poor human-machine motion compatibility,an equivalent series mechanism model that is highly matched with the actual bone structure of the human ankle joint is proposed and mapped into a parallel rehabilita-tion mechanism.The parallel rehabilitation mechanism has two virtual motion centers(VMCs),which can simulate the complex motion of the ankle joint,adapt to the individual differences of various patients,and can meet the reha-bilitation needs of both left and right feet of patients.Firstly,based on the motion properties and physiological structure of the human ankle joint,the mapping relationship between the rehabilitation mechanism and ankle joint is determined,and the series equivalent model of the ankle joint is established.According to the kinematic and con-straint properties of the ankle equivalent model,the configuration design of the parallel ankle rehabilitation robot is carried out.Secondly,according to the intersecting motion planes theory,the full-cycle mobility of the mechanism is proved,and the continuous axis of the mechanism is judged based on the constraint power and its derivative.Then,the kinematics of the parallel ankle rehabilitation robot is analyzed.Finally,based on the OpenSim biomechanical soft-ware,a human-machine coupling rehabilitation simulation model is established to evaluate the rehabilitation effect,which lays the foundation for the formulation of a rehabilitation strategy for the later prototype.