The effects of temperature and wavelength choice on in-situ dissolution test instrument of Cimetidine were studied. Absorbance (A)〈 1.0 is required when using a fiber-optic dissolution test system. The detection wa...The effects of temperature and wavelength choice on in-situ dissolution test instrument of Cimetidine were studied. Absorbance (A)〈 1.0 is required when using a fiber-optic dissolution test system. The detection wavelength of 2 (218 nm) was replaced by 244 nm to carry out this test. The absorbance of Cimetidine solution at different temperature showed an obvious change. Calibration of Cimetidine solution should be tested at the same temperature (37° C) with the test solution. A suitable wavelength with smaller tangent slope could be chosen for in-situ dissolution test of Cimetidine tablets.展开更多
基金the Xinjiang Uygur Autonomous Region Natural Science Fund (No.2011211A041) Xinjiang Uygur Autonomous Region Science and Technology Plan (No.200910107)
文摘The effects of temperature and wavelength choice on in-situ dissolution test instrument of Cimetidine were studied. Absorbance (A)〈 1.0 is required when using a fiber-optic dissolution test system. The detection wavelength of 2 (218 nm) was replaced by 244 nm to carry out this test. The absorbance of Cimetidine solution at different temperature showed an obvious change. Calibration of Cimetidine solution should be tested at the same temperature (37° C) with the test solution. A suitable wavelength with smaller tangent slope could be chosen for in-situ dissolution test of Cimetidine tablets.