A calibrating device for the Rogowski coil is developed, which can be used to calibrate the Rogowski coil having a partial response time within tens of nanoseconds. Its key component is a step current generator, which...A calibrating device for the Rogowski coil is developed, which can be used to calibrate the Rogowski coil having a partial response time within tens of nanoseconds. Its key component is a step current generator, which can generate the output with a rise time of less than 2 ns and a duration of larger than 300 ns. The step current generator is composed by a pulse forming line (PFL) and a pulse transmission line (PTL). A TEM (transverse electromagnetic mode) coaxial measurement unit is used as PTL, and the coil to be calibrated and the referenced standard Rogowski coil can be fixed in the unit. The effect of the dimensions of the TEM unit is discussed theoretically as well as experimentally.展开更多
文摘A calibrating device for the Rogowski coil is developed, which can be used to calibrate the Rogowski coil having a partial response time within tens of nanoseconds. Its key component is a step current generator, which can generate the output with a rise time of less than 2 ns and a duration of larger than 300 ns. The step current generator is composed by a pulse forming line (PFL) and a pulse transmission line (PTL). A TEM (transverse electromagnetic mode) coaxial measurement unit is used as PTL, and the coil to be calibrated and the referenced standard Rogowski coil can be fixed in the unit. The effect of the dimensions of the TEM unit is discussed theoretically as well as experimentally.