Performance test of the low-pressure thin window multi-wire chamber
1
作者 张俊伟 鲁辰桂 +5 位作者 段利敏 马龙 胡荣江 杨贺润 马朋 甘再国 《Chinese Physics C》 SCIE CAS CSCD 2014年第2期57-60,共4页
A flow gas low-pressure multi-wire proportional chamber(LPMWPC) with an active area of 180 mm× 80 mm has been developed for the flying time test of the recoil nuclei on super heavy nuclear experiments. The LPMW... A flow gas low-pressure multi-wire proportional chamber(LPMWPC) with an active area of 180 mm× 80 mm has been developed for the flying time test of the recoil nuclei on super heavy nuclear experiments. The LPMWPC detector can be operated in single as well as double step operational modes. In the case of double step operational mode with a high gas amplification factor, signals from α-particles reside well above the electronic noise. The gas leakage rate and time resolution obtained from the α239Pu source are shown and discussed at the condition of 3 mbar Isobutane gas. It was shown that the time resolution was better than 2.9 ns at the best work condition, and the detecting efciency was larger than 98% at the low energy α particles. So the LPMWPC is fit to measure the flying time in the super heavy nuclear fragments experiment. 展开更多
关键词 LPMWPC gas leakage rate time resolution detecting efciency