Objective: On the premise that the image quality meets the requirements of clinical diagnosis, we explored the methods to reduce the radiation dose of coronary artery imaging with Dual-Source CT (DSCT). Methods: We ra...Objective: On the premise that the image quality meets the requirements of clinical diagnosis, we explored the methods to reduce the radiation dose of coronary artery imaging with Dual-Source CT (DSCT). Methods: We randomly selected 200 patients with coronary heat disease (BIM 0.05). The average image noise in group A is (41.76 ± 7.98) HU, in group B the average image noise is (43.97 ± 3.88) HU, the dif- ference between the two groups was not statistically significant (P>0.05). The average CTDIvol of group A and B were (20.63 ± 2.24) mGy, (38.11 ± 10.69) mGy, respectively, then P <0.01. The average DLP of group A and B are (235.75 ± 28.64) mGycm and (492.59 ± 125.49) mGycm respectively, then P <0.01, the dif- ference of radiation dose had statistical significance (P<0.05). Conclusions: For coronary artery imaging with DSCT the heart electric pulse (AUTO) regulation technology can meet the diagnostic requirements and effectively reduce the radiation dose.展开更多
Objective: To compare and analyze the image quality and radiation dose of three scanning modes of dual-source CT coronary artery retrospectively, and to discuss the application value of TurboFlash coarse pitch scannin...Objective: To compare and analyze the image quality and radiation dose of three scanning modes of dual-source CT coronary artery retrospectively, and to discuss the application value of TurboFlash coarse pitch scanning mode. Methods: The imaging data of 100 patients who underwent CT coronary angiography (CCTA) using Siemens force CT retrospective gated triggering spiral scan (RES-SPIRAL), adaptive prospective gated triggering sequence scan (SEQ) and prospective coarse pitch scan (TurboFlash) retrospectively was collected. The image quality was evaluated by objective and subjective methods. The effective radiation dose of patients was compared and analyzed, and the indications of the three scanning modes were analyzed. The application value of dual-source CT TurboFlash coarse pitch scanning in coronary artery imaging was evaluated. Results: The results showed that the left main coronary artery, the right coronary artery and their tertiary branches could be clearly displayed in the three groups of images: the left anterior descending branch, the left circumflex branch, and their three-level branches. There was no statistical difference in subjective image quality among the three groups of pictures (P > 0.05). There was no statistical difference in objective evaluation indexes, such as CT value, SNR, CNR and Noise among the three groups (P > 0.05). The patient radiation dose results showed that the effective radiation dose ED of RES-SPIRA scan was (9.22 ± 1.33) mSv. The dose of SEQ was (2.88 ± 2.47) mSv, and the dose of TurboFlash was (0.51 ± 0.16) mSv. There was significant difference in comparison of the three groups (P 0.05). RES-spiral scanning had the highest radiation dose and TurboFlash coarse pitch scanning (TurboFlash) had the lowest radiation dose. Conclusion: TurboFlash coarse pitch scanning is low in dosage, fast in speed and wide in adaptability. It is especially suitable for the elderly, children, coma and other patients who cannot cooperate with breath-holding examination, as well as for the screening and examination of coronary artery diseases in asymptomatic population. Undoubtedly, it is a worthy method of heart coronary artery examination.展开更多
Objective To investigate the feasibility of acquiring the similar homogeneous enhancement using bolus-tracking techniques with shortened respiratory time in prospectively electrocardiogram-gated high-pitch spiral acqu...Objective To investigate the feasibility of acquiring the similar homogeneous enhancement using bolus-tracking techniques with shortened respiratory time in prospectively electrocardiogram-gated high-pitch spiral acquisition mode (Flash mode) coronary computed tomography angiography (CCTA) compared with test bolus technique. Methods One hundred and eighty-four consecutive patients with mean heart rate ≤65 beats per minute undergoing CCTA were prospectively included in this study. The patients were randomly divided into two groups. Patients in the group A (n=92) instructed to shorten respiratory time received CCTA using bolus-tracking technique with high-pitch spiral acquisition mode (Flash mode), while those in the group B (n=92) underwent CCTA with test bolus technique. The attenuation in the ascending aorta, image noise, contrast-to-noise ratio and radiation doses of the two groups were assessed. Results There were no significant differences in the mean attenuation values in the ascending aorta (483.18±59.07 HU vs. 498.7±83.51 HU, P=0.183), image noise (21.4±4.5 HU vs. 20.9±4.3 HU, P=0.414), contrast-to-noise ratio (12.1±4.2 vs. 13.8±5.1, P=0.31) between the groups A and B. There were no significant differences in the radiation dose of dynamic monitoring scans (0.056±0.026 mSv vs. 0.062±0.018 mSv, P=0.068) and radiation dose of angiography (0.94±0.07 mSv vs. 0.96±0.15 mSv,P=0.926) between the two groups, while 15 mL less contrast material volume was administered in the group A than the group B. Conclusion Bolus-tracking technique with shortened time of respiratory in Flash mode of dual-source CT yields the similar homogeneous enhancement with less contrast material in comparison to the test bolus technique.展开更多
Background Accurate evaluation of coronary artery,aortic valve annulus diameter (AVAD),and cardiac function in patients with aortic valve disease is of great significance for surgical strategy.In this study,we explo...Background Accurate evaluation of coronary artery,aortic valve annulus diameter (AVAD),and cardiac function in patients with aortic valve disease is of great significance for surgical strategy.In this study,we explored the preoperative evaluation of low-dose sequence (MinDose sequence) scan of dual-source CT (DSCT) for those patients.Methods Forty patients suspected for aortic valve disease (the experimental group) underwent MinDose sequence of DSCT to observe coronary artery,AVAD,and left ventricular ejection fraction (LVEF).Another 33 subjects suspected for coronary artery disease (the control group) underwent conventional retrospective electrocardiographically-gated sequence of DSCT.Two-dimensional transthoracic echocardiography (2D-TTE) and four-dimensional transthoracic echocardiography (4D-TTE) were applied in the experimental group to measure AVAD and LVEF and compared with MinDose-DSCT.Results There was a strong correlation between LVEFs measured by 2D-TTE and MinDose-DSCT (r=0.87,P <0.01),as well as between 4D-TTE and MinDose-DSCT (r=0.90,P <0.01).AVAD measured by MinDose-DSCT was in good agreement with corresponding measurements by 2D-TTE (r=0.90,P <0.01).The effective dose in the experimental group was 63.54% lower than that in the control group.Conclusions MinDose sequence of DSCT with a low radiation dose serving as a one-stop preoperative evaluation makes effective assessment of the coronary artery,AVAD,and LVEF for patients with aortic valve disease.展开更多
文摘Objective: On the premise that the image quality meets the requirements of clinical diagnosis, we explored the methods to reduce the radiation dose of coronary artery imaging with Dual-Source CT (DSCT). Methods: We randomly selected 200 patients with coronary heat disease (BIM 0.05). The average image noise in group A is (41.76 ± 7.98) HU, in group B the average image noise is (43.97 ± 3.88) HU, the dif- ference between the two groups was not statistically significant (P>0.05). The average CTDIvol of group A and B were (20.63 ± 2.24) mGy, (38.11 ± 10.69) mGy, respectively, then P <0.01. The average DLP of group A and B are (235.75 ± 28.64) mGycm and (492.59 ± 125.49) mGycm respectively, then P <0.01, the dif- ference of radiation dose had statistical significance (P<0.05). Conclusions: For coronary artery imaging with DSCT the heart electric pulse (AUTO) regulation technology can meet the diagnostic requirements and effectively reduce the radiation dose.
文摘Objective: To compare and analyze the image quality and radiation dose of three scanning modes of dual-source CT coronary artery retrospectively, and to discuss the application value of TurboFlash coarse pitch scanning mode. Methods: The imaging data of 100 patients who underwent CT coronary angiography (CCTA) using Siemens force CT retrospective gated triggering spiral scan (RES-SPIRAL), adaptive prospective gated triggering sequence scan (SEQ) and prospective coarse pitch scan (TurboFlash) retrospectively was collected. The image quality was evaluated by objective and subjective methods. The effective radiation dose of patients was compared and analyzed, and the indications of the three scanning modes were analyzed. The application value of dual-source CT TurboFlash coarse pitch scanning in coronary artery imaging was evaluated. Results: The results showed that the left main coronary artery, the right coronary artery and their tertiary branches could be clearly displayed in the three groups of images: the left anterior descending branch, the left circumflex branch, and their three-level branches. There was no statistical difference in subjective image quality among the three groups of pictures (P > 0.05). There was no statistical difference in objective evaluation indexes, such as CT value, SNR, CNR and Noise among the three groups (P > 0.05). The patient radiation dose results showed that the effective radiation dose ED of RES-SPIRA scan was (9.22 ± 1.33) mSv. The dose of SEQ was (2.88 ± 2.47) mSv, and the dose of TurboFlash was (0.51 ± 0.16) mSv. There was significant difference in comparison of the three groups (P 0.05). RES-spiral scanning had the highest radiation dose and TurboFlash coarse pitch scanning (TurboFlash) had the lowest radiation dose. Conclusion: TurboFlash coarse pitch scanning is low in dosage, fast in speed and wide in adaptability. It is especially suitable for the elderly, children, coma and other patients who cannot cooperate with breath-holding examination, as well as for the screening and examination of coronary artery diseases in asymptomatic population. Undoubtedly, it is a worthy method of heart coronary artery examination.
基金Supported by the Ministry of Science and Technology of Inner Mongolia, China (20110504)
文摘Objective To investigate the feasibility of acquiring the similar homogeneous enhancement using bolus-tracking techniques with shortened respiratory time in prospectively electrocardiogram-gated high-pitch spiral acquisition mode (Flash mode) coronary computed tomography angiography (CCTA) compared with test bolus technique. Methods One hundred and eighty-four consecutive patients with mean heart rate ≤65 beats per minute undergoing CCTA were prospectively included in this study. The patients were randomly divided into two groups. Patients in the group A (n=92) instructed to shorten respiratory time received CCTA using bolus-tracking technique with high-pitch spiral acquisition mode (Flash mode), while those in the group B (n=92) underwent CCTA with test bolus technique. The attenuation in the ascending aorta, image noise, contrast-to-noise ratio and radiation doses of the two groups were assessed. Results There were no significant differences in the mean attenuation values in the ascending aorta (483.18±59.07 HU vs. 498.7±83.51 HU, P=0.183), image noise (21.4±4.5 HU vs. 20.9±4.3 HU, P=0.414), contrast-to-noise ratio (12.1±4.2 vs. 13.8±5.1, P=0.31) between the groups A and B. There were no significant differences in the radiation dose of dynamic monitoring scans (0.056±0.026 mSv vs. 0.062±0.018 mSv, P=0.068) and radiation dose of angiography (0.94±0.07 mSv vs. 0.96±0.15 mSv,P=0.926) between the two groups, while 15 mL less contrast material volume was administered in the group A than the group B. Conclusion Bolus-tracking technique with shortened time of respiratory in Flash mode of dual-source CT yields the similar homogeneous enhancement with less contrast material in comparison to the test bolus technique.
文摘Background Accurate evaluation of coronary artery,aortic valve annulus diameter (AVAD),and cardiac function in patients with aortic valve disease is of great significance for surgical strategy.In this study,we explored the preoperative evaluation of low-dose sequence (MinDose sequence) scan of dual-source CT (DSCT) for those patients.Methods Forty patients suspected for aortic valve disease (the experimental group) underwent MinDose sequence of DSCT to observe coronary artery,AVAD,and left ventricular ejection fraction (LVEF).Another 33 subjects suspected for coronary artery disease (the control group) underwent conventional retrospective electrocardiographically-gated sequence of DSCT.Two-dimensional transthoracic echocardiography (2D-TTE) and four-dimensional transthoracic echocardiography (4D-TTE) were applied in the experimental group to measure AVAD and LVEF and compared with MinDose-DSCT.Results There was a strong correlation between LVEFs measured by 2D-TTE and MinDose-DSCT (r=0.87,P <0.01),as well as between 4D-TTE and MinDose-DSCT (r=0.90,P <0.01).AVAD measured by MinDose-DSCT was in good agreement with corresponding measurements by 2D-TTE (r=0.90,P <0.01).The effective dose in the experimental group was 63.54% lower than that in the control group.Conclusions MinDose sequence of DSCT with a low radiation dose serving as a one-stop preoperative evaluation makes effective assessment of the coronary artery,AVAD,and LVEF for patients with aortic valve disease.