A system is described in which a graphite furnace electrothermal vaporization device is employed for the introduction of microlitre liquid sample into an inductively coupled argon plasma. The technique provides a pico...A system is described in which a graphite furnace electrothermal vaporization device is employed for the introduction of microlitre liquid sample into an inductively coupled argon plasma. The technique provides a picogram detection limit and an adequate precision with a relative standard deviation of 4%. Mechanism of analyte condensation in transport process is explored. As an application, the technique combined with DDTC/CCl4 extraction is used to enrich and determine non- rare earth impurities in highly pure La2O3.展开更多
文摘A system is described in which a graphite furnace electrothermal vaporization device is employed for the introduction of microlitre liquid sample into an inductively coupled argon plasma. The technique provides a picogram detection limit and an adequate precision with a relative standard deviation of 4%. Mechanism of analyte condensation in transport process is explored. As an application, the technique combined with DDTC/CCl4 extraction is used to enrich and determine non- rare earth impurities in highly pure La2O3.