To study the detection method of trimethyltin chloride in urine.Extraction trimethyltin chloride in urine by ethyl acetate,and detection by capillary gas chromatography and mass spectrometry method (GC-MS).The linear ...To study the detection method of trimethyltin chloride in urine.Extraction trimethyltin chloride in urine by ethyl acetate,and detection by capillary gas chromatography and mass spectrometry method (GC-MS).The linear range of trimethyltin chloride in the method was 0.776-19.4 μg/mL,the correlation coefficient was 0.999 9,the lowest detection concentration was 0.01 mg/L.Relative standard deviation(RSD) was 4.4%-8.6%,the recovery rates were 99.0%-110.0%,sample could be save in -20℃ at least one half year.The method was applicable to detect trimethyltin chloride in urine.展开更多
文摘To study the detection method of trimethyltin chloride in urine.Extraction trimethyltin chloride in urine by ethyl acetate,and detection by capillary gas chromatography and mass spectrometry method (GC-MS).The linear range of trimethyltin chloride in the method was 0.776-19.4 μg/mL,the correlation coefficient was 0.999 9,the lowest detection concentration was 0.01 mg/L.Relative standard deviation(RSD) was 4.4%-8.6%,the recovery rates were 99.0%-110.0%,sample could be save in -20℃ at least one half year.The method was applicable to detect trimethyltin chloride in urine.