A rapid and practical method for direct detection of zidovudine in high performance anion exchange chromatography (HPAEC) has been developed with integrated pulsed amperometric detection (IPAD). Dionex AS18 (250m...A rapid and practical method for direct detection of zidovudine in high performance anion exchange chromatography (HPAEC) has been developed with integrated pulsed amperometric detection (IPAD). Dionex AS18 (250mm × 2mm) and AG18 (50 mm× 2 mm) columns and 11 mmol/L NaOH solution were used for separation. Multi-step potential waveform parameters were optimized to maximize the signal-to-noise ratio (S/N). Utilizing an optimized waveform, the repeatability (intra-day) precision and intermediate (inter-day) precision are obtained with relative standard deviation (RSD) of 1.88, 2.27, respectively. The limit of quantification (LOQ) and limit of detection (LOD) were found to be 9.70, 3.0 ng/mL, respectively, with correlation coefficients of 0.9992 over concentration range 0.01-10 μg/mL. The present method was successfully applied to the determination of zidovudine in human plasma. The recoveries of plasma sample spiked by 0.7 μg/mL, 2.7 μg/mL obtained were 95.3-101.5%, with relative standard deviation (RSD) of 2.54%, 2.21%, respectively.展开更多
基金supported by Zhejiang Provincial Natural Science Foundation of China(No.Y2110945)Zhejiang Provincial Analysis and Testing Foundation of China(No.2011C37065)
文摘A rapid and practical method for direct detection of zidovudine in high performance anion exchange chromatography (HPAEC) has been developed with integrated pulsed amperometric detection (IPAD). Dionex AS18 (250mm × 2mm) and AG18 (50 mm× 2 mm) columns and 11 mmol/L NaOH solution were used for separation. Multi-step potential waveform parameters were optimized to maximize the signal-to-noise ratio (S/N). Utilizing an optimized waveform, the repeatability (intra-day) precision and intermediate (inter-day) precision are obtained with relative standard deviation (RSD) of 1.88, 2.27, respectively. The limit of quantification (LOQ) and limit of detection (LOD) were found to be 9.70, 3.0 ng/mL, respectively, with correlation coefficients of 0.9992 over concentration range 0.01-10 μg/mL. The present method was successfully applied to the determination of zidovudine in human plasma. The recoveries of plasma sample spiked by 0.7 μg/mL, 2.7 μg/mL obtained were 95.3-101.5%, with relative standard deviation (RSD) of 2.54%, 2.21%, respectively.