Potentiometric Urea Biosensor Based on Ion-Selective Electrode by Different Immobilization Procedures
1
作者 Nien-Hsuan Chou Jung-Chuan Chou +1 位作者 Tai-Ping Sun Shen-Kan Hsiung 《稀有金属材料与工程》 SCIE EI CAS CSCD 北大核心 2006年第A03期309-312,共4页
Enzyme was immobilized on an ammonium ion-selective electrode by different methods.An ion-selective electrode is not completely ion-specific,and interfering ions react with the ion-selective electrode membrane,alterin... Enzyme was immobilized on an ammonium ion-selective electrode by different methods.An ion-selective electrode is not completely ion-specific,and interfering ions react with the ion-selective electrode membrane,altering the measured potential.Therefore,the characteristics of the effect of other ions on ammonium ion-selective electrode-based urea biosensors are considered.Based on the experimental results,the urea biosensor based on entrapment had a high response voltage of around 189 mV and fast response time of around 16 sec.Moreover,selectivity of the urea biosensor in different interfering ions was considered to elucidate the characteristics of ammonium ion-selective electrode-based biosensors. 展开更多
关键词 enzyme immobilization ammonium ion-selective electrode interfering ions urea biosensors SELECTIVITY