A novel method for the determination of nickel and palladium in environmental samples by low temperature ETV-ICP-OES with dimethylglyoxime(DMG) as both the extractant and chemical modifier has been developed. In thi...A novel method for the determination of nickel and palladium in environmental samples by low temperature ETV-ICP-OES with dimethylglyoxime(DMG) as both the extractant and chemical modifier has been developed. In this study, it was found that nickel and palladium can form complexes with dimethylglyoxime(0. 05%, mass fraction) at pH 6.0 and can be extracted into chloroform quantitatively. The complexes can be evaporated into plasma at a suita-ble temperature( 1400℃) for ICP-OES detection. Under the optimized conditions, the detection limits of nickel and palladium are 0.48 and 0. 40 ng/mL, respectively, while the RSD values are separately 5.0% and 3.1% (p = 50 ng/mL, n = 7). The proposed method was applied to the determination of the target analytes in environmental sam-ples with satisfactory results.展开更多
基金Supported by the National Natural Science Foundation of China(No.20575048).
文摘A novel method for the determination of nickel and palladium in environmental samples by low temperature ETV-ICP-OES with dimethylglyoxime(DMG) as both the extractant and chemical modifier has been developed. In this study, it was found that nickel and palladium can form complexes with dimethylglyoxime(0. 05%, mass fraction) at pH 6.0 and can be extracted into chloroform quantitatively. The complexes can be evaporated into plasma at a suita-ble temperature( 1400℃) for ICP-OES detection. Under the optimized conditions, the detection limits of nickel and palladium are 0.48 and 0. 40 ng/mL, respectively, while the RSD values are separately 5.0% and 3.1% (p = 50 ng/mL, n = 7). The proposed method was applied to the determination of the target analytes in environmental sam-ples with satisfactory results.