期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
Determination of Nickel and Palladium in Environmental Samples by Low Temperature ETV-ICP-OES Coupled with Liquid-liquid Extraction with Dimethylglyoxime as Both Extractant and Chemical Modifier 被引量:5
1
作者 PAN Liang QIN Yong-chao HU Bin JIANG Zu-cheng 《Chemical Research in Chinese Universities》 SCIE CAS CSCD 2007年第4期399-403,共5页
A novel method for the determination of nickel and palladium in environmental samples by low temperature ETV-ICP-OES with dimethylglyoxime(DMG) as both the extractant and chemical modifier has been developed. In thi... A novel method for the determination of nickel and palladium in environmental samples by low temperature ETV-ICP-OES with dimethylglyoxime(DMG) as both the extractant and chemical modifier has been developed. In this study, it was found that nickel and palladium can form complexes with dimethylglyoxime(0. 05%, mass fraction) at pH 6.0 and can be extracted into chloroform quantitatively. The complexes can be evaporated into plasma at a suita-ble temperature( 1400℃) for ICP-OES detection. Under the optimized conditions, the detection limits of nickel and palladium are 0.48 and 0. 40 ng/mL, respectively, while the RSD values are separately 5.0% and 3.1% (p = 50 ng/mL, n = 7). The proposed method was applied to the determination of the target analytes in environmental sam-ples with satisfactory results. 展开更多
关键词 low temperature electrothermal vaporization ICP-OES DIMETHYLGLYOXIME Chemical modifier Nickel and palladium
下载PDF
上一页 1 下一页 到第