A Rogowski coil is developed to detect the nanosecond pulse signals of the discharge current with a wide bandwidth of 800 kHz to 106 MHz and high sensitivity of 2.22 V/A. Performance tests show that the Rogowski coil ... A Rogowski coil is developed to detect the nanosecond pulse signals of the discharge current with a wide bandwidth of 800 kHz to 106 MHz and high sensitivity of 2.22 V/A. Performance tests show that the Rogowski coil has both excellent dynamic and static characteristics. Calibrating results and the comparison between the standard current shunt and the developed Rogowski coil for the measurement of nanosecond discharge pulses demonstrate that the developed Rogowski coil can reproduce the actual waveform of the discharge current accurately. 展开更多
关键词 Rogowski coil magnetic core MEASUREMENT nanosecond current pulses