The principle of the method for the BESIII TOF calibration using cosmic ray data without magnetic field are reported in this paper. After applying calibration constants, the single-end readout time resolution could re...The principle of the method for the BESIII TOF calibration using cosmic ray data without magnetic field are reported in this paper. After applying calibration constants, the single-end readout time resolution could reach about 150 ps, and the time resolution for one layer is achieved to be about 110 ps. The paper also described the extraction scheme for the event start time of cosmic events.展开更多
基金Supported by National Natural Science Foundation of China (10565001, 10647002, 10875138)
文摘The principle of the method for the BESIII TOF calibration using cosmic ray data without magnetic field are reported in this paper. After applying calibration constants, the single-end readout time resolution could reach about 150 ps, and the time resolution for one layer is achieved to be about 110 ps. The paper also described the extraction scheme for the event start time of cosmic events.