A method was proposed to determine boron trifluoride in boron trifluoride complex using fluoride ion selective electrode(ISE). Hydroxide was chosen to mask aluminum for the determination of 0.01—0.1 mol/L of fluoride...A method was proposed to determine boron trifluoride in boron trifluoride complex using fluoride ion selective electrode(ISE). Hydroxide was chosen to mask aluminum for the determination of 0.01—0.1 mol/L of fluoride. The simulation indicated that the permissible aluminum masked at a certain p H value was limited and hardly related to F-concentration and boric acid. It is better to control p H value below 11.5 and the aluminum concentration within 0.025 mol/L to minimize the interference of hydroxide to the fluoride ISE. The decomposition conditions of boron trifluoride by aluminum chloride were investigated. It is found that the F-detection ratio will approach 1.0 if the Al/F molar ratio is 0.3—0.7 and aluminum concentration is no more than 0.02 mol/L when heated at 80 ℃ for 10 min. In one word, hydroxide is quite fit to mask aluminum for samples which contain high content of fluoride and aluminum and the BF3 content can be successfully determined by this method.展开更多
A new chemically modified carbon paste electrode for cesium(I) ion determination based on potassium zinc hexacyan-oferrate (PZHCF) as an ionophore was prepared. The electrode exhibits a Nernstian response for Cs(I) io...A new chemically modified carbon paste electrode for cesium(I) ion determination based on potassium zinc hexacyan-oferrate (PZHCF) as an ionophore was prepared. The electrode exhibits a Nernstian response for Cs(I) ions over a wide concentration range from 1 × 10-6 to 1 × 10-1 mol·L-1 with a slope of 58 ± 0.5 mV·decade-1. It has a response time of about 35 s and can be used for a period of 3 months with good reproducibility. Detection limit obtained in the optimal conditions was 3 × 10-7 mol·L-1. The potentiometric response is independent of the pH of the solution in the pH range 4.0 - 8.0. The electrode possesses the advantages of low resistance, fast response over a variety of other cations. The proposed electrode is applied as a sensor for the determination of Cs(I) ion concentration in different samples solutions. The results showed a good correlation with the data obtained by atomic absorption spectrometric method.展开更多
In this paper is reported a new method of determining serum albumin with potentiometry. The silver ion selective electrode was used as working electrode and S. C. E as reference electrode. By controlling the ionic s...In this paper is reported a new method of determining serum albumin with potentiometry. The silver ion selective electrode was used as working electrode and S. C. E as reference electrode. By controlling the ionic strength, PH and temperature constant, a good linear relationship was obtained between the cell potential and albumin concentration. This method .possessed characters-of operating simplicity, rapidity and directness. The samples could be detected directly and the results were in accordance with the values detected by classic K method.展开更多
A PVC membrane enoxacin ion-selective electrode based on a needle-shaped inner reference electrode was prepared. A Ag/AgCl wire was used as the substrate of this electrode. It was previously coated with a thin sheet o...A PVC membrane enoxacin ion-selective electrode based on a needle-shaped inner reference electrode was prepared. A Ag/AgCl wire was used as the substrate of this electrode. It was previously coated with a thin sheet of urea-formaldehyde resin containing Cl - ions to form a needle-shaped inner reference electrode, then the inner reference electrode was coated with a thin sheet of a PVC membrane containing an enoxacin tetraphenylborate ion-pair complex. The influences of various ion-pair complexes, concentrations of the active components in the membrane and the plasticizers on the performance of the electrode were studied by orthogonal design. The linear response range of the electrode was 7.9×10 -5 -1.0×10 -2 mol/L. The detection limit was 2.0×10 -5 mol/L. The slope was 30.4 mV/decade(25 ℃). The electrode can be used for the potentiometric determination of enoxacin tablets directly. The average recovery was 100.4%, and the RSD was 0.9%. The results agreed with those determined by the method in Chinese Pharmacopoeia.展开更多
The construction principle of the system for determination of xanthate concentration,which consists of an Ionselective electrode(ISE) and a microprocessor based SDI analyzer,was described in the paper.The performances...The construction principle of the system for determination of xanthate concentration,which consists of an Ionselective electrode(ISE) and a microprocessor based SDI analyzer,was described in the paper.The performances of the developed system were discussed in details.The results both detected in solutions and in sulfide mineral flotation system were given; also the results by ultraviolet spectrometry were compared with those by the developed system.展开更多
The construction and performance characteristics of a novel Iron(III) membrane sensor based on (37-bis-[(diethoxy-thiophosphoryl)-oxy]-5,11,17,23,29,35-hexa-kis-(1,1-dimethylethyl)-calix[6]-arene-38,39,40,41,42-pentol...The construction and performance characteristics of a novel Iron(III) membrane sensor based on (37-bis-[(diethoxy-thiophosphoryl)-oxy]-5,11,17,23,29,35-hexa-kis-(1,1-dimethylethyl)-calix[6]-arene-38,39,40,41,42-pentol) (A) were reported in this paper. The sensor was prepared by incorporating compound (A) into a plasticized polyvinyl chloride PVC membrane. The electrode reveals a Nernstian behavior over a wide iron ion concentration range (1.0 × 10-2 - 1.0 × 10-4 mol·L-1), and relatively low detection limit (1.0 × 10-5 mol·L-1). The potentiometric response was independent on the pH of the solution in the range of 4 - 7. The electrode showed a very short response time (<5 s). It exhibited very good selectivity relative to a wide variety of metal cations. Correlation of selectivity coefficient values with ionic radii of interferent cations was discussed. The proposed iron- ISE was applied for analysis of iron containing samples with comparison to a recommended spectrophotometric method.展开更多
A new all-solid-state carbonate ion-selective electrode was fabricated via electrochemical and ion-exchange reactions with Ag nanoparticles and Ag2CO3-BaCO3 as ion-to-electron transducer and ion-selective layer, respe...A new all-solid-state carbonate ion-selective electrode was fabricated via electrochemical and ion-exchange reactions with Ag nanoparticles and Ag2CO3-BaCO3 as ion-to-electron transducer and ion-selective layer, respectively. This paper presented the detailed procedures of the construction of the electrode, the applications of the electrode in various experiments, and the corresponding chemical reaction principles involved. The fabricated electrode was characterized by means of scanning electron microscopy(SEM), energy dispersive spectrometer(EDS), and electrochemical analyses. The electrode exhibited stable linear responses to carbonate anions(10-5--10-1 mol/L) with average Slope of the curve being -26.56 mV per decade, and showed negligible responses to NO3, SO2-, and Sal- anions. The electrode was further used to measure the concentration of carbonate anions in boiler water. The measured concentration was lower than that determined via titration. Overall, the electrode exhibited the advantages of miniaturized size, robust construction, and compatibility with other equipment and thus could be ideally integrated into various sensor platforms to collect chemical data from industrial heat exchangers.展开更多
基金Supported by the Independent Innovation Fund of Tianjin University(No.1307)
文摘A method was proposed to determine boron trifluoride in boron trifluoride complex using fluoride ion selective electrode(ISE). Hydroxide was chosen to mask aluminum for the determination of 0.01—0.1 mol/L of fluoride. The simulation indicated that the permissible aluminum masked at a certain p H value was limited and hardly related to F-concentration and boric acid. It is better to control p H value below 11.5 and the aluminum concentration within 0.025 mol/L to minimize the interference of hydroxide to the fluoride ISE. The decomposition conditions of boron trifluoride by aluminum chloride were investigated. It is found that the F-detection ratio will approach 1.0 if the Al/F molar ratio is 0.3—0.7 and aluminum concentration is no more than 0.02 mol/L when heated at 80 ℃ for 10 min. In one word, hydroxide is quite fit to mask aluminum for samples which contain high content of fluoride and aluminum and the BF3 content can be successfully determined by this method.
文摘A new chemically modified carbon paste electrode for cesium(I) ion determination based on potassium zinc hexacyan-oferrate (PZHCF) as an ionophore was prepared. The electrode exhibits a Nernstian response for Cs(I) ions over a wide concentration range from 1 × 10-6 to 1 × 10-1 mol·L-1 with a slope of 58 ± 0.5 mV·decade-1. It has a response time of about 35 s and can be used for a period of 3 months with good reproducibility. Detection limit obtained in the optimal conditions was 3 × 10-7 mol·L-1. The potentiometric response is independent of the pH of the solution in the pH range 4.0 - 8.0. The electrode possesses the advantages of low resistance, fast response over a variety of other cations. The proposed electrode is applied as a sensor for the determination of Cs(I) ion concentration in different samples solutions. The results showed a good correlation with the data obtained by atomic absorption spectrometric method.
文摘In this paper is reported a new method of determining serum albumin with potentiometry. The silver ion selective electrode was used as working electrode and S. C. E as reference electrode. By controlling the ionic strength, PH and temperature constant, a good linear relationship was obtained between the cell potential and albumin concentration. This method .possessed characters-of operating simplicity, rapidity and directness. The samples could be detected directly and the results were in accordance with the values detected by classic K method.
文摘A PVC membrane enoxacin ion-selective electrode based on a needle-shaped inner reference electrode was prepared. A Ag/AgCl wire was used as the substrate of this electrode. It was previously coated with a thin sheet of urea-formaldehyde resin containing Cl - ions to form a needle-shaped inner reference electrode, then the inner reference electrode was coated with a thin sheet of a PVC membrane containing an enoxacin tetraphenylborate ion-pair complex. The influences of various ion-pair complexes, concentrations of the active components in the membrane and the plasticizers on the performance of the electrode were studied by orthogonal design. The linear response range of the electrode was 7.9×10 -5 -1.0×10 -2 mol/L. The detection limit was 2.0×10 -5 mol/L. The slope was 30.4 mV/decade(25 ℃). The electrode can be used for the potentiometric determination of enoxacin tablets directly. The average recovery was 100.4%, and the RSD was 0.9%. The results agreed with those determined by the method in Chinese Pharmacopoeia.
文摘The construction principle of the system for determination of xanthate concentration,which consists of an Ionselective electrode(ISE) and a microprocessor based SDI analyzer,was described in the paper.The performances of the developed system were discussed in details.The results both detected in solutions and in sulfide mineral flotation system were given; also the results by ultraviolet spectrometry were compared with those by the developed system.
文摘The construction and performance characteristics of a novel Iron(III) membrane sensor based on (37-bis-[(diethoxy-thiophosphoryl)-oxy]-5,11,17,23,29,35-hexa-kis-(1,1-dimethylethyl)-calix[6]-arene-38,39,40,41,42-pentol) (A) were reported in this paper. The sensor was prepared by incorporating compound (A) into a plasticized polyvinyl chloride PVC membrane. The electrode reveals a Nernstian behavior over a wide iron ion concentration range (1.0 × 10-2 - 1.0 × 10-4 mol·L-1), and relatively low detection limit (1.0 × 10-5 mol·L-1). The potentiometric response was independent on the pH of the solution in the range of 4 - 7. The electrode showed a very short response time (<5 s). It exhibited very good selectivity relative to a wide variety of metal cations. Correlation of selectivity coefficient values with ionic radii of interferent cations was discussed. The proposed iron- ISE was applied for analysis of iron containing samples with comparison to a recommended spectrophotometric method.
文摘A new all-solid-state carbonate ion-selective electrode was fabricated via electrochemical and ion-exchange reactions with Ag nanoparticles and Ag2CO3-BaCO3 as ion-to-electron transducer and ion-selective layer, respectively. This paper presented the detailed procedures of the construction of the electrode, the applications of the electrode in various experiments, and the corresponding chemical reaction principles involved. The fabricated electrode was characterized by means of scanning electron microscopy(SEM), energy dispersive spectrometer(EDS), and electrochemical analyses. The electrode exhibited stable linear responses to carbonate anions(10-5--10-1 mol/L) with average Slope of the curve being -26.56 mV per decade, and showed negligible responses to NO3, SO2-, and Sal- anions. The electrode was further used to measure the concentration of carbonate anions in boiler water. The measured concentration was lower than that determined via titration. Overall, the electrode exhibited the advantages of miniaturized size, robust construction, and compatibility with other equipment and thus could be ideally integrated into various sensor platforms to collect chemical data from industrial heat exchangers.