An ultralow power wireless intraocular pressure monitoring system
1
作者 刘德盟 梅年松 张钊锋 《Journal of Semiconductors》 EI CAS CSCD 2014年第10期177-181,共5页
This paper describes an ultralow power wireless intraocular pressure (IOP) monitoring system that is dedicated to sensing and transferring intraocular pressure of glaucoma patients. Our system is comprised of a capa... This paper describes an ultralow power wireless intraocular pressure (IOP) monitoring system that is dedicated to sensing and transferring intraocular pressure of glaucoma patients. Our system is comprised of a capac- itive pressure sensor, an application-specific integrated circuit, which is designed on the SMIC 180 nm process, and a dipole antenna. The system is wirelessly powered and demonstrates a power consumption of 7.56 μW at 1.24 V during continuous monitoring, a significant reduction in active power dissipation compared to existing work. The input RF sensitivity is -13 dBm. A significant reduction in input RF sensitivity results from the reduction of mis- match time of the ASK modulation caused by FM0 encoding. The system exhibits an average error of-4- 1.5 mmHg in measured pressure. Finally, a complete IOP system is demonstrated in the real biological environment, showing a successful reading of the pressure of an eye. 展开更多
关键词 IOP implanted medical pressure measurement RF powering ultralow power asic