The study of complex focussing acoustic field of a convex phased array probe used in ultrasonic diagnostic B-scanner
1
作者 LANG Yifu,BU Shuzhong, JIANG Nanxiang and CAI Chongcheng(Harbin Shipbuilding Engineering Institute ) 《Chinese Journal of Acoustics》 1991年第2期167-175,共9页
This paper reports a study of complex focussing acoustic fields of a convex phased array probe widely used in ultrasonic diagnostic B-scanner. The time delay necessary for electronic focussing and for geometric focuss... This paper reports a study of complex focussing acoustic fields of a convex phased array probe widely used in ultrasonic diagnostic B-scanner. The time delay necessary for electronic focussing and for geometric focussing of complex focussing is given. By means of Helmholtz's integral formula, the expressions of acoustic fields were obtained for the electronic focussing and complex focussing.With Simpson's numerical integral, a great deal of computations was carried out with an 86 / 330 computer. By means of using different parameters, an optimum focal range and beam width of a 4 sequences electronic and complex focussing acoustic system was selected. With an EUB-40 ultrasound diagnostic B-scanner produced by the Hitachi company, the lateral focussing acoustic field and lateral resolution in an echoic tank were photographed by latticing method and the experimental result is consistent with the theory. 展开更多
关键词 The study of complex focussing acoustic field of a convex phased array probe used in ultrasonic diagnostic B-scanner
原文传递
上一页 1 下一页 到第