A 2D neutron detector based on 3He convertor and MWPC with an active area of 200 mmx200 mm has been successfully designed and fabricated. The detector has been tested with Am/Be neutron source and with collimated neut...A 2D neutron detector based on 3He convertor and MWPC with an active area of 200 mmx200 mm has been successfully designed and fabricated. The detector has been tested with Am/Be neutron source and with collimated neutron beam with the wavelength of λ=1.37A. The best spatial resolution of 1.18 mm (FWHM) and good linearity were obtained. This is in good agreement with theoretical calculations.展开更多
基金supported by the National Natural Science Foundation of China(Grant No.11127508)
文摘A 2D neutron detector based on 3He convertor and MWPC with an active area of 200 mmx200 mm has been successfully designed and fabricated. The detector has been tested with Am/Be neutron source and with collimated neutron beam with the wavelength of λ=1.37A. The best spatial resolution of 1.18 mm (FWHM) and good linearity were obtained. This is in good agreement with theoretical calculations.