We developed a chiral HPLC method for the separation and analysis ofnaftopidil enantiomers. The two enantiomers of naflopidil were separated using a Chiralpak AD-H (250 mm×4.6 mm, 5 μm) column and monitored at...We developed a chiral HPLC method for the separation and analysis ofnaftopidil enantiomers. The two enantiomers of naflopidil were separated using a Chiralpak AD-H (250 mm×4.6 mm, 5 μm) column and monitored at the wavelength of 283 nm. The isocratic mobile phase consisting of hexane-isopropanol-diethylamine (85:15:0.1, v/v/v) was pumped at a flow rate of 1.0 mL/min. Under these chromatographic conditions, R-naflopidil and S-naftopidil were well separated and had good linearity in the ranges of 0.78-50 μg/mL (r = 0.9999) and 0.84-54 μg/mL (r = 0.9998), respectively. The relative standard deviations (RSD) of intra- and inter-day assays were no more than 0.5% and 0.7%, respectively. This improved method for the separation and quantitative determination of naflopidil enantiomers can be used for the quality control of synthesized naflopidil product.展开更多
基金Guangzhou Science and Technology Projects of Apecial Areas of Innovative Medicines (Grant No.2006Z2-E4011)2006 Guangdong Province Technology Projects (Grant No. 2006B35501003)
文摘We developed a chiral HPLC method for the separation and analysis ofnaftopidil enantiomers. The two enantiomers of naflopidil were separated using a Chiralpak AD-H (250 mm×4.6 mm, 5 μm) column and monitored at the wavelength of 283 nm. The isocratic mobile phase consisting of hexane-isopropanol-diethylamine (85:15:0.1, v/v/v) was pumped at a flow rate of 1.0 mL/min. Under these chromatographic conditions, R-naflopidil and S-naftopidil were well separated and had good linearity in the ranges of 0.78-50 μg/mL (r = 0.9999) and 0.84-54 μg/mL (r = 0.9998), respectively. The relative standard deviations (RSD) of intra- and inter-day assays were no more than 0.5% and 0.7%, respectively. This improved method for the separation and quantitative determination of naflopidil enantiomers can be used for the quality control of synthesized naflopidil product.