Objective: To study the effect of surface configuration on the corrosion resistance and bacterium accumulation of pure Titanium(Ti) when used in oral environment. Methods: Six edentulous volunteers with healthy oral m...Objective: To study the effect of surface configuration on the corrosion resistance and bacterium accumulation of pure Titanium(Ti) when used in oral environment. Methods: Six edentulous volunteers with healthy oral mucosa participated in an in vivo study. Two kinds of pure Ti testing pieces with plane and wrinkly surface were fixed in the polished surface of upper complete dentures. After 6-month wearing, dynamic polarization curves were traced with electrochemical method and the amount and species of bacterium adhered on pure Ti were examined. Results: ① Ep and Ip of specimen in oral cavity was higher than that left in air, which meant corrosion resistance falling. Compared to plane one, Ecorr of wrinkly specimen was more positive and Ep and Ip were more higher, so its corrosion resistance reduced. ② Individual difference put a significant influence on amount of bacterium adhered on pure Titanium, but had no relation to species of bacterium. To the same patient, wrinkly samples collected more bacterium than plane ones, and exhibited G - coccus beside G + coccus. Conclusion: From the perspective of corrosion behavior and benefit to periodontal tissue, wrinkly surface should not be adopted when pure Ti prosthesis is made,especially on connector part of denture.展开更多
文摘Objective: To study the effect of surface configuration on the corrosion resistance and bacterium accumulation of pure Titanium(Ti) when used in oral environment. Methods: Six edentulous volunteers with healthy oral mucosa participated in an in vivo study. Two kinds of pure Ti testing pieces with plane and wrinkly surface were fixed in the polished surface of upper complete dentures. After 6-month wearing, dynamic polarization curves were traced with electrochemical method and the amount and species of bacterium adhered on pure Ti were examined. Results: ① Ep and Ip of specimen in oral cavity was higher than that left in air, which meant corrosion resistance falling. Compared to plane one, Ecorr of wrinkly specimen was more positive and Ep and Ip were more higher, so its corrosion resistance reduced. ② Individual difference put a significant influence on amount of bacterium adhered on pure Titanium, but had no relation to species of bacterium. To the same patient, wrinkly samples collected more bacterium than plane ones, and exhibited G - coccus beside G + coccus. Conclusion: From the perspective of corrosion behavior and benefit to periodontal tissue, wrinkly surface should not be adopted when pure Ti prosthesis is made,especially on connector part of denture.