In order to determine the low interfacial tension,a spinning drop interfacial tensiometer was constructed.The apparatus was calibrated by measuring the interfacial tension of some low molecular weight systems and poly...In order to determine the low interfacial tension,a spinning drop interfacial tensiometer was constructed.The apparatus was calibrated by measuring the interfacial tension of some low molecular weight systems and polyethylene glycol /dextran aqueous two-phase systems,with the determination error of 3.The interfacial tensions of PEG1000/Na2SO4,PEG/MgSO4 and PEG1000/(NH4)2SO4 aqueous two-phase systems Were determined.It is found that the salt type has strong effect on the interfacial tension.展开更多
文摘In order to determine the low interfacial tension,a spinning drop interfacial tensiometer was constructed.The apparatus was calibrated by measuring the interfacial tension of some low molecular weight systems and polyethylene glycol /dextran aqueous two-phase systems,with the determination error of 3.The interfacial tensions of PEG1000/Na2SO4,PEG/MgSO4 and PEG1000/(NH4)2SO4 aqueous two-phase systems Were determined.It is found that the salt type has strong effect on the interfacial tension.