In the practical measurement of signal to noise ratio(SNR)of analog-to-digital converters(ADCs)by using fast Fourier transformation(FFT)method,the non-coherent sampling is inevitable,leading to spectral leakage which ...In the practical measurement of signal to noise ratio(SNR)of analog-to-digital converters(ADCs)by using fast Fourier transformation(FFT)method,the non-coherent sampling is inevitable,leading to spectral leakage which in turn affects the calculation accuracy and final measurement results.In this paper,a new method based on the Blackman-Harris windowed triple-spectrum-line interpolation is presented for the measurement of ADCs SNR by FFT.The simulation platform is built based on MATLAB and the behavioral dynamic models of the high-speed ADC products of Analog Devices Inc.(ADI)are simulated.The simulation results show that,even in the case of the maximum non-coherent degree,the SNR error is less than0.23dB and reaches the testing standards provided by ADI,confirming that the proposed method is effective for suppressing the spectral leakage effects and improving the SNR test accuracy.展开更多
基金Summit of the Six Top Talents Program of Jiangsu Province(No.2013-DZXX-027)Fundamental Research Funds for the Central Universities(Nos.JUSRP51510,JUSRP51323B)Graduate Student Innovation Program for Universities of Jiangsu Province(Nos.SJLX16_0500,KYLX16_0776,SJCX17_0510)
文摘In the practical measurement of signal to noise ratio(SNR)of analog-to-digital converters(ADCs)by using fast Fourier transformation(FFT)method,the non-coherent sampling is inevitable,leading to spectral leakage which in turn affects the calculation accuracy and final measurement results.In this paper,a new method based on the Blackman-Harris windowed triple-spectrum-line interpolation is presented for the measurement of ADCs SNR by FFT.The simulation platform is built based on MATLAB and the behavioral dynamic models of the high-speed ADC products of Analog Devices Inc.(ADI)are simulated.The simulation results show that,even in the case of the maximum non-coherent degree,the SNR error is less than0.23dB and reaches the testing standards provided by ADI,confirming that the proposed method is effective for suppressing the spectral leakage effects and improving the SNR test accuracy.