Cadmium has been determined by β-correction spectrophotometry with cadion,p-nitrobenzenediazoaminoaz-obenzone,and a non-ionic surfactant,tuiton X-100.The real absorbance of a Cd-cadion chelate in the colored solution...Cadmium has been determined by β-correction spectrophotometry with cadion,p-nitrobenzenediazoaminoaz-obenzone,and a non-ionic surfactant,tuiton X-100.The real absorbance of a Cd-cadion chelate in the colored solution can be accurately determined and the complex-ratio of cadion with Cd(II) has been worked out to be 2.Beer's law is obeyed over the concentration range of 0-0.20mg/1 cadmium and the detec-tion limit for cadmium is only 0.003mg/1.Satisfactory experimental results are presented with respect to the determination of trace cadmium in wastewaters.展开更多
文摘Cadmium has been determined by β-correction spectrophotometry with cadion,p-nitrobenzenediazoaminoaz-obenzone,and a non-ionic surfactant,tuiton X-100.The real absorbance of a Cd-cadion chelate in the colored solution can be accurately determined and the complex-ratio of cadion with Cd(II) has been worked out to be 2.Beer's law is obeyed over the concentration range of 0-0.20mg/1 cadmium and the detec-tion limit for cadmium is only 0.003mg/1.Satisfactory experimental results are presented with respect to the determination of trace cadmium in wastewaters.