A preliminary study of small-mass radiocarbon sample measurement at Xi'an-AMS
1
作者 付云翀 周卫健 +5 位作者 杜花 程鹏 赵晓雷 刘起 卢雪峰 赵稳年 《Chinese Physics C》 SCIE CAS CSCD 2015年第3期45-52,共8页
To meet the measurement demands on small-mass radiocarbon (carbon content at 10.6 g level) which are becoming increasingly significant, Xi'an-AMS has made improvements to the existing method of sample loading and h... To meet the measurement demands on small-mass radiocarbon (carbon content at 10.6 g level) which are becoming increasingly significant, Xi'an-AMS has made improvements to the existing method of sample loading and has upgraded the Cs sputter ion source from the original SO-110 model. In order to study the feasibility of small-mass samples in Xi'an-AMS and evaluate the radiocarbon sample preparation ability using existing routine systems of H2/Fe and Zn/Fe, the small-mass samples prepared by four different methods are tested. They are the mass division method, mass dilution method, H2/Fe reduction method and Zn/Fe reduction method. The results show that carbon mass above 25 μg can be prepared using the existing Zn/Fe system, but no less than 100 μg is required using the existing H2/Fe system, which can be improved. This indicates Xi'an-AMS are now able to analyze small-mass radiocarbon samples. 展开更多
关键词 ams small-mass radiocarbon sample ion source