In this report, the results of preliminary research on L-alanine/ESR electron dosimeters are described. They include the designs of modelling tools and experimental capsules, the preparation and irradiation of very th...In this report, the results of preliminary research on L-alanine/ESR electron dosimeters are described. They include the designs of modelling tools and experimental capsules, the preparation and irradiation of very thin dosimeters, the measurement of ESR signal spectrum and the calibration method. The physical, chemical and dosimetric properties of the alanine free radical dosimeters, such as the smallest thickness, average density, lowest detectable limit, repeatability of the method, reproducibility resulted in directivity and location in resonance cavity, linearity regions as well as relative scattering of the response to identical doses, are summarized.展开更多
基金This project was supported by IAEA under the research contact No. 4236/RB/R1
文摘In this report, the results of preliminary research on L-alanine/ESR electron dosimeters are described. They include the designs of modelling tools and experimental capsules, the preparation and irradiation of very thin dosimeters, the measurement of ESR signal spectrum and the calibration method. The physical, chemical and dosimetric properties of the alanine free radical dosimeters, such as the smallest thickness, average density, lowest detectable limit, repeatability of the method, reproducibility resulted in directivity and location in resonance cavity, linearity regions as well as relative scattering of the response to identical doses, are summarized.