Employing arbitrary Lagrangian-Eulerian (ALE) finite element method, this poper studies the opening and closing process of a St. Jude medical valve through a two-dimensional model of the mechanical valve-blood interac...Employing arbitrary Lagrangian-Eulerian (ALE) finite element method, this poper studies the opening and closing process of a St. Jude medical valve through a two-dimensional model of the mechanical valve-blood interaction in which the valve is regarded as a rigid body rotating around a fixed point, and foe blood is simplified as viscous incompressible Newtonian fluid. The numerical analysis of the opening and closing behaviour of as St. Jude valve suggested that: 1. The whole opening and closing process of an artificial mechanical valve is consisted of four phases: (1) Opening phase; (2) Opening maintenance phase; (3) Closing phase; (4) Closing maintenance phase. 2. The St. Jude medical valve closes with prominent regurgitat which results in water-hammer effect. 3. During the opening and closing process of the St. Jude valve,high shear stresses occur in the middle region of the two leaflets and on the valve ring. The present model has made a breakthrough on the coupling computational analysis considering the interactive movement of the valve and blood.展开更多
Using arbitrary Lagrangian-Eulerian(ALE)finite element method,this paper made a comparative study of the opening and closing behaviour of a downstream directional valve(DDM)and a St.Jude medical valve(SJM)through a tw...Using arbitrary Lagrangian-Eulerian(ALE)finite element method,this paper made a comparative study of the opening and closing behaviour of a downstream directional valve(DDM)and a St.Jude medical valve(SJM)through a two dimensional model of mechanical valve-blood interaction in which the valve is considered as a rigid body rotating around a fixed point,and the blood is simplified as viscous incompressible fluid It's concluded that:(1)Compared with SJM valve, DDM valve opens faster and closes the more gently.(2)The peak badk-flow-flow of DDM is smaller than that of SJM.The present investigation shows that being a better analogue of natural valve,DDM has a brighter potential on its durability than SJM.展开更多
文摘Employing arbitrary Lagrangian-Eulerian (ALE) finite element method, this poper studies the opening and closing process of a St. Jude medical valve through a two-dimensional model of the mechanical valve-blood interaction in which the valve is regarded as a rigid body rotating around a fixed point, and foe blood is simplified as viscous incompressible Newtonian fluid. The numerical analysis of the opening and closing behaviour of as St. Jude valve suggested that: 1. The whole opening and closing process of an artificial mechanical valve is consisted of four phases: (1) Opening phase; (2) Opening maintenance phase; (3) Closing phase; (4) Closing maintenance phase. 2. The St. Jude medical valve closes with prominent regurgitat which results in water-hammer effect. 3. During the opening and closing process of the St. Jude valve,high shear stresses occur in the middle region of the two leaflets and on the valve ring. The present model has made a breakthrough on the coupling computational analysis considering the interactive movement of the valve and blood.
文摘Using arbitrary Lagrangian-Eulerian(ALE)finite element method,this paper made a comparative study of the opening and closing behaviour of a downstream directional valve(DDM)and a St.Jude medical valve(SJM)through a two dimensional model of mechanical valve-blood interaction in which the valve is considered as a rigid body rotating around a fixed point,and the blood is simplified as viscous incompressible fluid It's concluded that:(1)Compared with SJM valve, DDM valve opens faster and closes the more gently.(2)The peak badk-flow-flow of DDM is smaller than that of SJM.The present investigation shows that being a better analogue of natural valve,DDM has a brighter potential on its durability than SJM.