Performance Calibration of Articulated Arm Coordinate Measuring Machine 被引量:5
1
作者 Unai Mutilba Gorka Kortaberria Aitor Olarra Antonio Gutierrez 《Journal of Mechanics Engineering and Automation》 2013年第10期602-608,共7页
This paper presents the method for the performance calibration of AACMM (articulated arm coordinate measuring machines) according to ASME B89.4.22 Standard. The growing use of this class of measurement equipment has... This paper presents the method for the performance calibration of AACMM (articulated arm coordinate measuring machines) according to ASME B89.4.22 Standard. The growing use of this class of measurement equipment has been accompanied by an absence of authorized laboratories to provide calibration certificates for its performance. Due to ASME B89.4.22 and VD12617-9 are nowadays the unique standards in the field of AACMM verification, IK4 Tekniker has compared both of them in order to develop internal test procedures to yield reliable performance calibration results. As a result, IK4 Tekniker has been recognized by the Spanish Accreditation Body (ENAC) in the field of AACMM calibration. Internal test procedures and uncertainty evaluation analysis have been developed as well as ENAC certificated reference test equipments have been acquired to ensure a suitable AACMM calibration process. 展开更多
关键词 AACMM performance calibration ASME b89.4.22 ENAC UNCERTAINTY