Based on the first ionization chamber (IC) prototype, the structure, working gas component and electrode material of the IC are improved. The test of the improved IC shows that the plateau length is about 2000 V, th...Based on the first ionization chamber (IC) prototype, the structure, working gas component and electrode material of the IC are improved. The test of the improved IC shows that the plateau length is about 2000 V, the plateau slope is less than 0.2%/100 V, the sensitivity is 19.6 pA/rad·h 1, the up-limitation of the linearity can be up to 3.6×105 rad/h, and the applied voltage can be operated to 3500 V. The test results show that the performance of the improved IC meets the requirements of the beam loss monitor.展开更多
文摘Based on the first ionization chamber (IC) prototype, the structure, working gas component and electrode material of the IC are improved. The test of the improved IC shows that the plateau length is about 2000 V, the plateau slope is less than 0.2%/100 V, the sensitivity is 19.6 pA/rad·h 1, the up-limitation of the linearity can be up to 3.6×105 rad/h, and the applied voltage can be operated to 3500 V. The test results show that the performance of the improved IC meets the requirements of the beam loss monitor.