Polydimethylsiloxane Microchip Capillary Electrophoresis for Determination of Cholesterol
1
作者 Nipapan Ruecha Weena Siangproh Orawon Chailapakul 《Journal of Electronic Science and Technology of China》 2010年第1期82-86,共5页
Microchip capillary electrophoresis (MCE) has significant impact on diagnostic testing. One of the most importance in clinical analysis test is the determination of cholesterol level in blood, because its increase i... Microchip capillary electrophoresis (MCE) has significant impact on diagnostic testing. One of the most importance in clinical analysis test is the determination of cholesterol level in blood, because its increase is associated with coronary heart disease which is a major cause of death world over. Polydimethylsiloxane (PDMS) was material used to make a MCE for determination of cholesterol. Hydrogen peroxide was generated from the oxidation of cholesterol with cholesterol oxidase (ChOx) and was detected electrochemically at a downstream gold (Au) wire electrode by amperometry. Various parameters, such as the detection potential, concentration of running buffer, pH of running buffer, separation voltage, injection time, and concentration of ChOx, were studied. The migration time of cholesterol is less than 100s and the calibration curve is linear from 50 mg.dL-1 to 250 mg-dL-1 with the coefficient of determination (R2) of 0.9955. Therefore, this proposed assay is very rapid and sensitive for the detection of cholesterol. 展开更多
关键词 cholesterol cholesterol oxidase microchip capillary electrophoresis polydimethylsiloxane.
下载PDF
上一页 1 下一页 到第