A prototype of a two-dimensional position sensitive X-ray detector was designed and constructed for small angle X-ray scattering experiments at BSFR (Beijing Synchrotron Radiation Facility). The detector is based on...A prototype of a two-dimensional position sensitive X-ray detector was designed and constructed for small angle X-ray scattering experiments at BSFR (Beijing Synchrotron Radiation Facility). The detector is based on MWPC with cathode strip readout, and has a sensitive area of 200 mm×200 mm. The spatial resolution (FWHM) of about 210 μm along the anode wire direction was obtained from the ^55Fe X-ray test of the detector.展开更多
文摘A prototype of a two-dimensional position sensitive X-ray detector was designed and constructed for small angle X-ray scattering experiments at BSFR (Beijing Synchrotron Radiation Facility). The detector is based on MWPC with cathode strip readout, and has a sensitive area of 200 mm×200 mm. The spatial resolution (FWHM) of about 210 μm along the anode wire direction was obtained from the ^55Fe X-ray test of the detector.