目的了解GyneFixIN宫内节育器的临床使用效果。方法随机给396例健康已婚育龄妇女放置3种不同宫内节育器,其中GyneFix IN IUD135例,TCu380A130例,对照组TCu220C131例,在放置后的3、12个月进行随访观察。结果使用12个月时3组IUD妊娠率分别...目的了解GyneFixIN宫内节育器的临床使用效果。方法随机给396例健康已婚育龄妇女放置3种不同宫内节育器,其中GyneFix IN IUD135例,TCu380A130例,对照组TCu220C131例,在放置后的3、12个月进行随访观察。结果使用12个月时3组IUD妊娠率分别为0.7%,0.8%,2.3%,GyneFixINIUD和TCu380A明显低于TCu220C(P<0.05);脱落率分别为0%,6.9%,8.4%,GyneFix IN明显低于TCu380A和TCu220C;因症取出率分别为0%,5.4%,5.3%,GyneFixIN明显低于其他2组差异有显著性意义(P<0.01)。结论GyneFixIN具有脱落率低,不良反应少,避孕效果好,是目前较理想的宫内节育器。展开更多
To analyze the effect of two types of IUDs, TCu380A and GyneFix on the positive rate of Chlamydia Trahmatis (CT). Methods The TCu380A and GyneFix IUDs were compared in a randomized was for the one year and two year...To analyze the effect of two types of IUDs, TCu380A and GyneFix on the positive rate of Chlamydia Trahmatis (CT). Methods The TCu380A and GyneFix IUDs were compared in a randomized was for the one year and two year positive rate of Chlamydia Trahmatis (CT) and with the control respectively. Results The one year positive rate of CT antigens was 5.63% of TCu380A and two year was 4.92%; the one year positive rate of CT antigens was 4.62% and two year was 5.08% of GyneFix. There was no significant difference in the positive rate of CT antigen between the TCu380A IUD, and GyneFix IUDs groups, while there were significant differences between the TCu380A IUD, GyneFix IUD and the controls (15.18%) respectively. Conclusion Both IUDs provide highly effective protection against CT infection.展开更多
文摘目的了解GyneFixIN宫内节育器的临床使用效果。方法随机给396例健康已婚育龄妇女放置3种不同宫内节育器,其中GyneFix IN IUD135例,TCu380A130例,对照组TCu220C131例,在放置后的3、12个月进行随访观察。结果使用12个月时3组IUD妊娠率分别为0.7%,0.8%,2.3%,GyneFixINIUD和TCu380A明显低于TCu220C(P<0.05);脱落率分别为0%,6.9%,8.4%,GyneFix IN明显低于TCu380A和TCu220C;因症取出率分别为0%,5.4%,5.3%,GyneFixIN明显低于其他2组差异有显著性意义(P<0.01)。结论GyneFixIN具有脱落率低,不良反应少,避孕效果好,是目前较理想的宫内节育器。
文摘To analyze the effect of two types of IUDs, TCu380A and GyneFix on the positive rate of Chlamydia Trahmatis (CT). Methods The TCu380A and GyneFix IUDs were compared in a randomized was for the one year and two year positive rate of Chlamydia Trahmatis (CT) and with the control respectively. Results The one year positive rate of CT antigens was 5.63% of TCu380A and two year was 4.92%; the one year positive rate of CT antigens was 4.62% and two year was 5.08% of GyneFix. There was no significant difference in the positive rate of CT antigen between the TCu380A IUD, and GyneFix IUDs groups, while there were significant differences between the TCu380A IUD, GyneFix IUD and the controls (15.18%) respectively. Conclusion Both IUDs provide highly effective protection against CT infection.