In this study, we fabricated a Cerenkov fiber-optic dosimeter (CFOD) without any scintillator to measure Cerenkov radiation signals owing to gamma-rays. The relative depth dose (RDD) distributions of Ir-192 HDR brachy...In this study, we fabricated a Cerenkov fiber-optic dosimeter (CFOD) without any scintillator to measure Cerenkov radiation signals owing to gamma-rays. The relative depth dose (RDD) distributions of Ir-192 HDR brachytherapy source were obtained by using the CFOD based on a subtraction method and the RDD curve was compared with the simulation result of Monte Carlo N-particle extended transport code (MCNPX). Finally, we demonstrated that the CFOD can be used to measure real-time dose information for HDR brachytherapy.展开更多
文摘In this study, we fabricated a Cerenkov fiber-optic dosimeter (CFOD) without any scintillator to measure Cerenkov radiation signals owing to gamma-rays. The relative depth dose (RDD) distributions of Ir-192 HDR brachytherapy source were obtained by using the CFOD based on a subtraction method and the RDD curve was compared with the simulation result of Monte Carlo N-particle extended transport code (MCNPX). Finally, we demonstrated that the CFOD can be used to measure real-time dose information for HDR brachytherapy.