Cathodic adsorptive stripping voltammetry of an anti-emetic agent Granisetron in pharmaceutical formulation and biological matrix
1
作者 Rajeev Jain Ramkishor Sharma 《Journal of Pharmaceutical Analysis》 SCIE CAS 2012年第6期443-449,共7页
Granisetron showed one well-defined reduction peak at Hanging Mercury Drop Electrode (HMDE) in the potential range from -1.3 to -1.5 V due to reduction of C=N bond. Solid-phase extraction technique was employed for ... Granisetron showed one well-defined reduction peak at Hanging Mercury Drop Electrode (HMDE) in the potential range from -1.3 to -1.5 V due to reduction of C=N bond. Solid-phase extraction technique was employed for extraction of Granisetron from spiked human plasma. Granisetron showed peak current enhancement of 4.45% at square-wave voltammetry and 5.33% at cyclic voltammetry as compared with the non stripping techniques. The proposed voltammetric method allowed quantification of Granisetron in pharmaceutical formulation over the target concentration range of 5-200 ng/mL with detection limit 13.63 ng/mL, whereas in human plasma 50-225 ng/mL with detection limit 11.75 ng/mL. 展开更多
关键词 GRANISETRON Human plasma Solid-phase extraction Pharmaceuticalformulation VOLTAMMETRY hanging mercury dropelectrode