期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
A SPHERICAL GRAPHITE IONIZATION CHAMBER
1
作者 陈丽姝 《Nuclear Science and Techniques》 SCIE CAS CSCD 1992年第3期167-170,共4页
This paper describes an exposure ratemeter made of graphite ion chamber. The sensitive volume is 14 cm^3. The wall is 5 mm in thickness. The DC amplifier with a high input impedance consists of several stage different... This paper describes an exposure ratemeter made of graphite ion chamber. The sensitive volume is 14 cm^3. The wall is 5 mm in thickness. The DC amplifier with a high input impedance consists of several stage differential amplifications, and is used for measurement of the ionization currents. Six ranges are able to cover a wide range of exposure rates. The full scale of the maximum range is 75 C/(kg·h). The lowest measurable limit is 5 mC/(kg·h). The zero drift within 8 h is less than 5% of each full scale. This instrument provides with a relatively convenient method for the measurement. The main advantages of the ionization chamber over other methods lie in rapidity and accuracy, and the change in radiation field with time can be indicated directly. It is particularly successful in acting as a routine dosimeter under high exposure rates. 展开更多
关键词 high dose exposure ratemeter gamma dose spherical ionization chamber cavity ionization chamber