期刊文献+
共找到3篇文章
< 1 >
每页显示 20 50 100
Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma
1
作者 Shunsuke Nakakura Ryo Asaoka +5 位作者 Etsuko Terao Yuki Nagata Yasuko Fukuma Satomi Oogi Miku Shiraishi Yoshiaki Kiuchi 《Eye and Vision》 SCIE CSCD 2021年第1期244-252,共9页
Background:This study investigated the agreement between a new rebound tonometer,IC200,and IcarePRO and Goldmann applanation tonometry(GAT).Methods:This was a prospective cross-sectional study.We measured the intraocu... Background:This study investigated the agreement between a new rebound tonometer,IC200,and IcarePRO and Goldmann applanation tonometry(GAT).Methods:This was a prospective cross-sectional study.We measured the intraocular pressure(IOP)in 145 eyes of 145 glaucoma patients in the sitting position using GAT,IcarePRO,and IC200.IcarePRO and IC200 measurements were also obtained in the supine position.IC200 measurement was performed using two modes:single six(IC200-single)and automatic(IC200-continuous)six-measurements mode.Results:All tonometers provided high reproducibility in both positions(all intraclass correlation coefficients>0.90),although it was highest with GAT,followed by IC200-continuous and IC200-single and then IcarePRO.In the sitting position,the mean(±SD)IOPs of GAT,IcarePRO,IC200-single,and IC200-continuous were 14.5±2.9 mmHg,13.3±3.2 mmHg,11.6±3.2 mmHg,and 11.5±3.2 mmHg,respectively.IOPs measured with IcarePRO or IC200 were significantly lower than those with GAT,particularly in patients with low IOP.IOPs measured with all tonometers were significantly elevated in the supine position as compared with the sitting position,but this difference was significantly greater with IC200-single and IC200-continuous compared with IcarePRO.IOP elevation was significant in eyes without bleb versus those with bleb,but this finding was not observed when IOP was measured with IcarePRO.The IOPs of the single and continuous modes of IC200 were interchangeable in both positions.Conclusions:GAT,IcarePRO,and IC200 had sufficiently high reproducibility,but measurements with IcarePRO may not be accurate in the supine position.Elevation of IOP in the supine position,especially in eyes with bleb,was more sensitively captured with IC200 than with IcarePRO. 展开更多
关键词 GLAUCOMA ic200 IcarePRO Rebound tonometer Goldmann applanation tonometer
原文传递
小区庭院道路照明设计
2
作者 朱英莉 王兆滨 《林业科技情报》 2004年第2期55-55,共1页
本文介绍了小区道路照明设计特点,并就设计中遇到的照明控制问题进行了探讨。
关键词 住宅小区 庭院道路 照明设计 照明控制
下载PDF
SCHMID内话系统常用配置应用分析
3
作者 闫飞龙 《科技创新导报》 2019年第1期159-160,171,共3页
民航空管系统中的地空通信设备种类繁多,配置方法和使用方式也各有所长,本文以瑞士SCHMID公司生产的ICS200/60型地空通信系统为例,从基础入手,介绍该系统配置/监控子系统的实际应用,并通过工作中的配置操作案例,逐步深入,结合日常工作... 民航空管系统中的地空通信设备种类繁多,配置方法和使用方式也各有所长,本文以瑞士SCHMID公司生产的ICS200/60型地空通信系统为例,从基础入手,介绍该系统配置/监控子系统的实际应用,并通过工作中的配置操作案例,逐步深入,结合日常工作中分析解决问题时总结出来的经验,对维护过程中经常使用的一些配置方法和关键参数进行了分析。 展开更多
关键词 ICS200/60 配置 应用 分析
下载PDF
上一页 1 下一页 到第
使用帮助 返回顶部