To study the effect of guide vane outlet angle on pump performance and impeller radial force in an axial-flow heart pump, guide vane outlet angle/34 is considered to be 20°, 15°, 10°, 5° and 3°...To study the effect of guide vane outlet angle on pump performance and impeller radial force in an axial-flow heart pump, guide vane outlet angle/34 is considered to be 20°, 15°, 10°, 5° and 3° respectively. Based on ANSYS Fluent, numerical results of pump head and efficiency are validated by exper/ment results, in which Xanthan gum solutions are used with concentration of 0.06 wt. % as working fluid. Then, the effects of/34 on pump performance and impeller radial force are discussed, the errors of head and efficiency between test and simulation are within 5%. The results a/so indicate that the pump performance and efficiency are much better than those of other angles when guide vane outlet angle/34 is 10°, and the maximum variations in head and efficiency are 1.9% and 2.2%, respectively. With/34 increasing, the pulsation of radial force decreases firstly and then increases, when/34 is 10°, the minimum pulsation is 0. 0392N, which is about 80% of the maximum pulsation amplitude.展开更多
基金Supported by the National Natural Science Foundation of China(No.51574161)the Education Science Project of Young and Middle-aged Teachers of Universities in Fujian Province(No.JZ160396)
文摘To study the effect of guide vane outlet angle on pump performance and impeller radial force in an axial-flow heart pump, guide vane outlet angle/34 is considered to be 20°, 15°, 10°, 5° and 3° respectively. Based on ANSYS Fluent, numerical results of pump head and efficiency are validated by exper/ment results, in which Xanthan gum solutions are used with concentration of 0.06 wt. % as working fluid. Then, the effects of/34 on pump performance and impeller radial force are discussed, the errors of head and efficiency between test and simulation are within 5%. The results a/so indicate that the pump performance and efficiency are much better than those of other angles when guide vane outlet angle/34 is 10°, and the maximum variations in head and efficiency are 1.9% and 2.2%, respectively. With/34 increasing, the pulsation of radial force decreases firstly and then increases, when/34 is 10°, the minimum pulsation is 0. 0392N, which is about 80% of the maximum pulsation amplitude.