Purpose: To present an application of the anisotropic diffusion (AD) method to improve the accuracy of the functional images of perfusion parameters such as cerebral blood flow (CBF), cerebral blood volume (CBV) and m...Purpose: To present an application of the anisotropic diffusion (AD) method to improve the accuracy of the functional images of perfusion parameters such as cerebral blood flow (CBF), cerebral blood volume (CBV) and mean transit time (MTT) generated from cerebral CT perfusion studies using multi-detector row CT (MDCT). Materials and Methods: Continuous scans (1 sec/rotation ×60 sec) consisting of four 5-mm-thick contiguous slices were acquired after an intravenous injection of iodinated contrast material in 6 patients with cerebrovascular disease using an MDCT scanner with a tube voltage of 80 kVp and a tube current of 200 mA. New image data were generated by thinning out the above original images at an interval of 2 sec or 3 sec. The thinned-out images were then interpolated by linear interpolation to generate the same number of images as originally acquired. The CBF, CBV and MTT images were generated using deconvolution analysis based on singular value decomposition. Results: When using the AD method, the correlation coefficient between the MTT values obtained from the original and thinned-out images was significantly improved. Furthermore, the coefficients of variation of the CBF, CBV and MTT values in the white matter significantly decreased as compared to not using the AD method. Conclusion: Our results suggest that the AD method is useful for improving the accuracy of the functional images of perfusion parameters and for reducing radiation exposure in cerebral CT perfusion studies using MDCT.展开更多
文摘Purpose: To present an application of the anisotropic diffusion (AD) method to improve the accuracy of the functional images of perfusion parameters such as cerebral blood flow (CBF), cerebral blood volume (CBV) and mean transit time (MTT) generated from cerebral CT perfusion studies using multi-detector row CT (MDCT). Materials and Methods: Continuous scans (1 sec/rotation ×60 sec) consisting of four 5-mm-thick contiguous slices were acquired after an intravenous injection of iodinated contrast material in 6 patients with cerebrovascular disease using an MDCT scanner with a tube voltage of 80 kVp and a tube current of 200 mA. New image data were generated by thinning out the above original images at an interval of 2 sec or 3 sec. The thinned-out images were then interpolated by linear interpolation to generate the same number of images as originally acquired. The CBF, CBV and MTT images were generated using deconvolution analysis based on singular value decomposition. Results: When using the AD method, the correlation coefficient between the MTT values obtained from the original and thinned-out images was significantly improved. Furthermore, the coefficients of variation of the CBF, CBV and MTT values in the white matter significantly decreased as compared to not using the AD method. Conclusion: Our results suggest that the AD method is useful for improving the accuracy of the functional images of perfusion parameters and for reducing radiation exposure in cerebral CT perfusion studies using MDCT.