To accelerate the scan speed and improve the image quality, a new type of CT configuration, "double-source double-multislice spiral CT" (DSDMS-CT), which is based on two sets of single-source multislice spir...To accelerate the scan speed and improve the image quality, a new type of CT configuration, "double-source double-multislice spiral CT" (DSDMS-CT), which is based on two sets of single-source multislice spiral CT was proposed with a special reconstruction algorithm. Simulation results using the fan-beam filtered backprojection algorithm with a special interpolation method were presented for both single-source multislice spiral CT and DSDMS-CT. The results of new CT model show that it scans faster than the traditional spiral CT and has a better slice sensitivity profile (SSP) with larger pitch value.展开更多
基金The National Natural Science Foundation ofChina (No30570511)
文摘To accelerate the scan speed and improve the image quality, a new type of CT configuration, "double-source double-multislice spiral CT" (DSDMS-CT), which is based on two sets of single-source multislice spiral CT was proposed with a special reconstruction algorithm. Simulation results using the fan-beam filtered backprojection algorithm with a special interpolation method were presented for both single-source multislice spiral CT and DSDMS-CT. The results of new CT model show that it scans faster than the traditional spiral CT and has a better slice sensitivity profile (SSP) with larger pitch value.