In this paper full-energy peak (photopeak) efficiency and photopeak attenuation coefficient of 3'' × 3'' NaI(Tl) well-type scintillation detector were calculated using gamma-rayisotropic radiating...In this paper full-energy peak (photopeak) efficiency and photopeak attenuation coefficient of 3'' × 3'' NaI(Tl) well-type scintillation detector were calculated using gamma-rayisotropic radiating point sources (with photon energy: 0.245, 0.344, 0.662, 0.779, 0.964, 1.1732, 1.333 and 1.408 MeV) placed outside the detector well. These energies were obtained from <sup>152</sup>Eu, <sup>137</sup>Cs and <sup>60</sup>Co. The relations between the full energy peak efficiency and photopeak attenuation coefficients, were plotted vs. photon energy at different sources to detector distance, and it found that the full energy peak efficiency decreased by increasing the distance between the source and the detector.展开更多
文摘In this paper full-energy peak (photopeak) efficiency and photopeak attenuation coefficient of 3'' × 3'' NaI(Tl) well-type scintillation detector were calculated using gamma-rayisotropic radiating point sources (with photon energy: 0.245, 0.344, 0.662, 0.779, 0.964, 1.1732, 1.333 and 1.408 MeV) placed outside the detector well. These energies were obtained from <sup>152</sup>Eu, <sup>137</sup>Cs and <sup>60</sup>Co. The relations between the full energy peak efficiency and photopeak attenuation coefficients, were plotted vs. photon energy at different sources to detector distance, and it found that the full energy peak efficiency decreased by increasing the distance between the source and the detector.