Performance tests of a portable thyroid radioiodine meter
1
作者 Zeshu Li Zhijie Yang +3 位作者 Fei Tuo Shuo Wang Qianqian Zhou Juncheng Liang 《Radiation Medicine and Protection》 CSCD 2024年第1期53-58,共6页
Objective:To evaluate the effectiveness of a portable thyroid radioiodine meter in on-site monitoring to enhance the monitoring precision of radioiodine inhaled by nuclear medicine staff and patients.Methods:A portabl... Objective:To evaluate the effectiveness of a portable thyroid radioiodine meter in on-site monitoring to enhance the monitoring precision of radioiodine inhaled by nuclear medicine staff and patients.Methods:A portable thyroid radioiodine meter was developed to measure the 131I activity in the thyroid.Performance tests of the portable thyroid radioiodine meter were conducted using calibration sources and testing methodologies adhering to national metrological standards.Results:The thyroid radioiodine meter exhibited integral non-linearity,8-h measurement stability,measurement linearity,and an energy resolution of 0.42%,1.05%,1.56%,and 7.11%,respectively for ^(137)Cs at 661.66 keV.Additionally,its minimum detectable activity for ^(131)I at 364.49 keV was approximately 22 Bq(10 min).Conclusion:The developed thyroid radioiodine meter features elevated performance and great potential for application in nuclear medicine and monitoring of responses to nuclear or radiological emergencies. 展开更多
关键词 Thyroid radioiodine meter Performance test nuclear medicine nuclear or radiological emergency