An investigation using Monte Carlo simulation on a minitype reference radiation(MRR) for the calibration of gamma personal dosimeters is reported. The distributions of dose rate and scattering gamma spectrum are the m...An investigation using Monte Carlo simulation on a minitype reference radiation(MRR) for the calibration of gamma personal dosimeters is reported. The distributions of dose rate and scattering gamma spectrum are the main simulation objects with the variable physical structures of MRR and the dosimeters as parameters that are to be calibrated. Further, the influences on the reference radiation caused by these parameters are analyzed in detail.This work provides a theoretical basis for better understanding of MRR used for calibration of gamma personal dosimeters. This analysis can help in the development of a calibration technology for such tools based on MRR.展开更多
基金supported by the National Natural Science Foundation of China(Nos.11805111 and 11525521)the Fundamental Research Funds for the Central Universities(No.2018CDGFGD0008)
文摘An investigation using Monte Carlo simulation on a minitype reference radiation(MRR) for the calibration of gamma personal dosimeters is reported. The distributions of dose rate and scattering gamma spectrum are the main simulation objects with the variable physical structures of MRR and the dosimeters as parameters that are to be calibrated. Further, the influences on the reference radiation caused by these parameters are analyzed in detail.This work provides a theoretical basis for better understanding of MRR used for calibration of gamma personal dosimeters. This analysis can help in the development of a calibration technology for such tools based on MRR.