Two time-domain reflectometry (TDR) systems and a new impedance measuring instrument, Thetaprobe,which are based on determination of soil dielectric constant, were used to measure water content of clayeyred soil to er...Two time-domain reflectometry (TDR) systems and a new impedance measuring instrument, Thetaprobe,which are based on determination of soil dielectric constant, were used to measure water content of clayeyred soil to eraluate the accuracy of these instruments. The results indicated that these instruments shouldbe carefUlly re-calibrated before being applied in clayey red soil. With a new calibration curve fed into one ofthe TDR systems tested, nase system, the measured data compared well with tho8e by standard oven-dryingmethod.展开更多
文摘Two time-domain reflectometry (TDR) systems and a new impedance measuring instrument, Thetaprobe,which are based on determination of soil dielectric constant, were used to measure water content of clayeyred soil to eraluate the accuracy of these instruments. The results indicated that these instruments shouldbe carefUlly re-calibrated before being applied in clayey red soil. With a new calibration curve fed into one ofthe TDR systems tested, nase system, the measured data compared well with tho8e by standard oven-dryingmethod.