The quality of radiation therapy depends on the ability to maximize the tumor control probability while minimizing the normal tissue complication probability.Both of these two quantities are directly related to the ac...The quality of radiation therapy depends on the ability to maximize the tumor control probability while minimizing the normal tissue complication probability.Both of these two quantities are directly related to the accuracy of dose distributions calculated by treatment planning systems.The commonly used dose calculation algorithms in the treatment planning systems are reviewed in this work.The accuracy comparisons among these algorithms are illustrated by summarizing the highly cited research papers on this topic.Further,the correlation between the algorithms and tumor control probability/normal tissue complication probability values are manifested by several recent studies from different groups.All the cases demonstrate that dose calculation algorithms play a vital role in radiation therapy.展开更多
Objective: The aim of this work was to quantify the extent of set-up errors to conduct a quality assurance (QA) aspect of treatment delivery, verification of the treatment field's position on different days using ...Objective: The aim of this work was to quantify the extent of set-up errors to conduct a quality assurance (QA) aspect of treatment delivery, verification of the treatment field's position on different days using electronic portal. Methods: This study was carried out on 12 patients, treated for pelvis tumor; and total of 240 images obtained by electronic portal image device (EPID) were analyzed. The EPIs acquire using EPID attached to the Siemens linear accelerator. The anatomy match- ing software (Theraview) was used and displacement in two dimensions were noted for each treatment field to study patient setup errors. Results: The percentages of mean deviations less than 5 mm in X direction were 65% & 92%, from 5-10 mm were 31% & 19% and more than 10 mm were 11% & 9% forNP and lateral direction respectively. The percentages of mean deviations less than 5 mm in Y direction were 65% & 63%, from 5-10 mm were 33% & 28% and more than 10 mm were 22% & 29%. The mean deviations in 2D-vector errors were 〈 5 mm in 47% and 46%, 5-10 mm in 36% and 37% and 〉 10 mm in 37% and 37% of images in the NP and lateral direction respectively. Conclusion: The results revealed that the ranges of set up errors are immobilization method to improve reproducibility. The observed variations were not within the limits..展开更多
基金Supported by In part,under a grant with the Pennsylvania Department of Health
文摘The quality of radiation therapy depends on the ability to maximize the tumor control probability while minimizing the normal tissue complication probability.Both of these two quantities are directly related to the accuracy of dose distributions calculated by treatment planning systems.The commonly used dose calculation algorithms in the treatment planning systems are reviewed in this work.The accuracy comparisons among these algorithms are illustrated by summarizing the highly cited research papers on this topic.Further,the correlation between the algorithms and tumor control probability/normal tissue complication probability values are manifested by several recent studies from different groups.All the cases demonstrate that dose calculation algorithms play a vital role in radiation therapy.
文摘Objective: The aim of this work was to quantify the extent of set-up errors to conduct a quality assurance (QA) aspect of treatment delivery, verification of the treatment field's position on different days using electronic portal. Methods: This study was carried out on 12 patients, treated for pelvis tumor; and total of 240 images obtained by electronic portal image device (EPID) were analyzed. The EPIs acquire using EPID attached to the Siemens linear accelerator. The anatomy match- ing software (Theraview) was used and displacement in two dimensions were noted for each treatment field to study patient setup errors. Results: The percentages of mean deviations less than 5 mm in X direction were 65% & 92%, from 5-10 mm were 31% & 19% and more than 10 mm were 11% & 9% forNP and lateral direction respectively. The percentages of mean deviations less than 5 mm in Y direction were 65% & 63%, from 5-10 mm were 33% & 28% and more than 10 mm were 22% & 29%. The mean deviations in 2D-vector errors were 〈 5 mm in 47% and 46%, 5-10 mm in 36% and 37% and 〉 10 mm in 37% and 37% of images in the NP and lateral direction respectively. Conclusion: The results revealed that the ranges of set up errors are immobilization method to improve reproducibility. The observed variations were not within the limits..