Abstract: This paper describes the development and characterization of a two-temperature (2-T), constant pressure humidity generator It relies on the saturation of a stream of gas flowing over a water surface maint...Abstract: This paper describes the development and characterization of a two-temperature (2-T), constant pressure humidity generator It relies on the saturation of a stream of gas flowing over a water surface maintained at constant, well-known, temperature. It was built in order to improve the uncertainties of the dew-point temperature and humidity scales realization at the National Institute for Standard (NIS) in the dew-point range from -50 ℃ to + 10 ℃. Several experiments were carried out in the above mentioned range in order to characterize the generator. Characterization comprised studies of its saturator efficiency, temperature stability and a comparison with a calibrated chilled-mirror hygrometer. The results of the efficiency tests showed good performance of the generator as described below. For uncertainty of measurements, a thorough analysis was also described representing estimations of contributions for all the sources that affecting measurements.展开更多
文摘Abstract: This paper describes the development and characterization of a two-temperature (2-T), constant pressure humidity generator It relies on the saturation of a stream of gas flowing over a water surface maintained at constant, well-known, temperature. It was built in order to improve the uncertainties of the dew-point temperature and humidity scales realization at the National Institute for Standard (NIS) in the dew-point range from -50 ℃ to + 10 ℃. Several experiments were carried out in the above mentioned range in order to characterize the generator. Characterization comprised studies of its saturator efficiency, temperature stability and a comparison with a calibrated chilled-mirror hygrometer. The results of the efficiency tests showed good performance of the generator as described below. For uncertainty of measurements, a thorough analysis was also described representing estimations of contributions for all the sources that affecting measurements.