Effects of temperature and wavelength choice on in-situ dissolution test of Cimetidine tablets 被引量:1
1
作者 Xin-Xia Li Yan Wang +6 位作者 Ping-Ping Xu Qi-Zhou Zhang Kun Nie Xu Hu Bin Kong Li Li Jian Chen 《Journal of Pharmaceutical Analysis》 SCIE CAS 2013年第1期71-74,共4页
The effects of temperature and wavelength choice on in-situ dissolution test instrument of Cimetidine were studied. Absorbance (A)〈 1.0 is required when using a fiber-optic dissolution test system. The detection wa... The effects of temperature and wavelength choice on in-situ dissolution test instrument of Cimetidine were studied. Absorbance (A)〈 1.0 is required when using a fiber-optic dissolution test system. The detection wavelength of 2 (218 nm) was replaced by 244 nm to carry out this test. The absorbance of Cimetidine solution at different temperature showed an obvious change. Calibration of Cimetidine solution should be tested at the same temperature (37° C) with the test solution. A suitable wavelength with smaller tangent slope could be chosen for in-situ dissolution test of Cimetidine tablets. 展开更多
关键词 cimetidine tablets Drug dissolution test In-situ dissolution test UV-VIS