Lessons learned in delayed identification of a misplaced electrode array in the vestibule
1
作者 Huey Ting Diong Yuhan Wong +1 位作者 Shermaine Png Yeow Seng Raymond Ngo 《Journal of Otology》 CAS CSCD 2024年第4期241-246,共6页
To report a case of cochlear implantation with a misplaced electrode array in the vestibule and the causes for the delay in identification.A 23-year-old male with left single-sided deafness underwent cochlear implanta... To report a case of cochlear implantation with a misplaced electrode array in the vestibule and the causes for the delay in identification.A 23-year-old male with left single-sided deafness underwent cochlear implantation.The intraoperative assessment did not reveal any major red flags of electrode array misplacement.He did not display any vestibular symptoms postoperatively but showed poor speech performance,even though the aided tone audiometry revealed good sound detection thresholds.High-resolution computed tomography(HRCT)showed that the entire perimodiolar electrode array was situated within the vestibule,and a revision surgery was conducted.Retrospective analysis of the neural response telemetry(NRT)revealed subtle differences in responses between the misplaced and correctly placed electrode arrays.Unlike previously reported cases,the patient did not display vestibular symptoms despite the misplacement of the electrode in the vestibule due to existing weakness in otolithic function.Further investigation is warranted when a motivated patient with normal inner ear anatomy does not show benefit with the cochlear implant post-operatively. 展开更多
关键词 cochlear implantation electrode misplacement VESTIBULE Poor CI outcome neural response telemetry
下载PDF
Nucleus人工耳蜗电极阻抗测试的研究 被引量:4
2
作者 王超 王丽娜 +1 位作者 肖玉丽 金德均 《中国现代医学杂志》 CAS CSCD 北大核心 2006年第1期92-94,共3页
目的探讨NucleusCI24M与NucleusCI24R(CS)型人工耳蜗电极阻抗的区别。方法应用神经反应遥测技术(NRT)监测16例NucleusCI24M与10例NucleusCI24R(CS)型人工耳蜗电极的阻抗。结果所有电极阻抗值正常,两种类型的人工耳蜗电极阻抗差异存在显... 目的探讨NucleusCI24M与NucleusCI24R(CS)型人工耳蜗电极阻抗的区别。方法应用神经反应遥测技术(NRT)监测16例NucleusCI24M与10例NucleusCI24R(CS)型人工耳蜗电极的阻抗。结果所有电极阻抗值正常,两种类型的人工耳蜗电极阻抗差异存在显著性,并且人工耳蜗电极植入后,其阻抗会发生规律性变化,在术中测试电极阻抗数值存在从蜗底到蜗顶逐渐上升的现象。结论NucleusCI24M与Nu-cleusCI24R(CS)的电极阻抗存在明显差异,电极植入后,其阻抗值随时间发生的规律性变化与机体生理活动有关。在相同条件下NucleusCI24R(CS)比NucleusCI24M更加节能有效。 展开更多
关键词 人工耳蜗 神经反应遥测技术(NRT) 电极阻抗
下载PDF
内耳畸形小儿人工耳蜗植入术后神经反应遥测值的研究 被引量:2
3
作者 乔晓峰 王东 雒星梅 《中国耳鼻咽喉头颈外科》 CSCD 2014年第1期29-30,共2页
目的探讨小儿内耳畸形人工耳蜗植入手术后,植入体不同电极神经反应遥测(neural response telemetry,NRT)的变化特点及规律。方法对36例澳大利亚Cochlear Nucleus 24型内耳畸形小儿人工耳蜗植入手术患儿,测试手术中及手术后1年NRT值,并... 目的探讨小儿内耳畸形人工耳蜗植入手术后,植入体不同电极神经反应遥测(neural response telemetry,NRT)的变化特点及规律。方法对36例澳大利亚Cochlear Nucleus 24型内耳畸形小儿人工耳蜗植入手术患儿,测试手术中及手术后1年NRT值,并分析其特点及变化规律。结果患儿术中及开机后1年内同一时间高、中、低频段电极NRT值有显著差异(P<0.05),低频值较低,高频值较高,总的趋势是术中至术后1年NRT值逐渐增高。结论内耳畸形小儿人工耳蜗植入手术后NRT值随时间逐渐增高,高、中、低频之间差异大,故调机时要注意各频率之间差异,选取高、中、低频代表性电极测试,以使患者语言感知达到最佳状态。 展开更多
关键词 迷路 畸形 儿童 耳蜗植入术 神经反应遥测
下载PDF
人工耳蜗植入术儿童多频稳态诱发反应与NRT值和T值的相关性研究 被引量:3
4
作者 乔晓峰 白伟红 +1 位作者 王东 李同丽 《中国中西医结合耳鼻咽喉科杂志》 2015年第5期321-322,334,共3页
目的探讨人工耳蜗植入术儿童多频稳态诱发反应(multiple auditory steady state responses,ASSR)阈值与手术后1年神经反应遥测(neural response telemetry,NRT)阈值及行为反应阈值(T-level,T级)的相关性。方法将128例澳大利亚Freedom人... 目的探讨人工耳蜗植入术儿童多频稳态诱发反应(multiple auditory steady state responses,ASSR)阈值与手术后1年神经反应遥测(neural response telemetry,NRT)阈值及行为反应阈值(T-level,T级)的相关性。方法将128例澳大利亚Freedom人工耳蜗植入患儿,依据术前多频稳态诱发电位高、中、低频有无残余听力,分别分为有残余听力组和无残余听力组,测定其术后1年NRT阈值及行为反应阈值,并分析对比其差异性。结果高、中、低频有残余听力组患儿术后1年NRT阈值及行为反应阈值均低于无残余听力组患儿,两者统计学比较有显著差异(P<0.01)。结论术前依据多频稳态检测结果分组,两组术后神经反应遥测阈值及行为反应阈值具有显著差异,故人工耳蜗植入术患儿术前进行多频稳态检测,可以预估其术后听力反应,且可作为术耳选择参考条件之一。 展开更多
关键词 人工耳蜗植入 神经反应遥测 多频稳态 儿童
下载PDF
人工耳蜗植入术术中检测的临床应用
5
作者 王一红 李瑞玉 +3 位作者 陈洪 黄少鹏 蒋刈 陈平 《福建医药杂志》 CAS 2008年第2期3-5,共3页
目的分析电极阻抗测试和神经反应遥测(NRT)技术在人工耳蜗手术中的应用,评估人工耳蜗植入术术中电极阻抗测试、NRT的临床意义,探讨术中检测的方式及注意事项。方法对12例植入Nucleus24型人工耳蜗的患者于术中及术后1个月进行电极阻抗测... 目的分析电极阻抗测试和神经反应遥测(NRT)技术在人工耳蜗手术中的应用,评估人工耳蜗植入术术中电极阻抗测试、NRT的临床意义,探讨术中检测的方式及注意事项。方法对12例植入Nucleus24型人工耳蜗的患者于术中及术后1个月进行电极阻抗测试、NRT,并比较术中、术后ECAP阈值的差异。结果91.7%(11/12)的患者术中电极阻抗测试正常;8.3%(1/12)的患者其中1个电极出现开路。91.7%(11/12)的患者记录到良好的NRT反应波形;8.3%(1/12)的患者术中NRT测试未获得分化良好的ECAP。术中、术后ECAP阈值具有显著的相关性。结论电极阻抗测试可以及时了解电极的工作状况。NRT技术可以有效地应用于Nucleus人工耳蜗植入患者的术中检测,同时协助开机时不能进行心理物理测试的儿童完成映射图的调试。 展开更多
关键词 入工耳蜗 手术 电极 听神经复合动作电位 神经反应遥测
下载PDF
小儿正常耳蜗与内耳畸形人工耳蜗植入术后电极阻抗及NRT值对比研究 被引量:6
6
作者 乔晓峰 王东 《临床耳鼻咽喉头颈外科杂志》 CAS 北大核心 2013年第22期1231-1233,共3页
目的:探讨小儿正常耳蜗与内耳畸形人工耳蜗植入手术后,植入体电极阻抗值及神经反应遥测(NRT)值的变化特点及规律。方法:将88例澳大利亚Cochlear Nucleus 24型人工耳蜗植入手术患儿分为正常耳蜗组与内耳畸形耳蜗组,测试手术中及手术后1... 目的:探讨小儿正常耳蜗与内耳畸形人工耳蜗植入手术后,植入体电极阻抗值及神经反应遥测(NRT)值的变化特点及规律。方法:将88例澳大利亚Cochlear Nucleus 24型人工耳蜗植入手术患儿分为正常耳蜗组与内耳畸形耳蜗组,测试手术中及手术后1年阻抗值及NRT值,并分析对比。结果:两组患者阻抗值差异有统计学意义(P>0.05),其阻抗变化趋势相同。开机3个月内阻抗值变化明显,其后趋于稳定。两组患者NRT值差异有统计学意义(P<0.05),其NRT值变化趋势相同。总的趋势是低频值较低,高频值较高,术中至术后1年NRT值逐渐增高。结论:电极阻抗值及NRT值开机3个月内变化明显,故需多次调机,以使患者语言感知达到最佳状态。 展开更多