Determination of Timer Error and Evaluation of Its Effect on Dose for OB6, GammaBeam X200 and X-Ray Irradiators at the Secondary Standard Dosimetry Laboratory in Nigeria
1
作者 Olumide Olaife Akerele Samuel Mofolorunsho Oyeyemi +3 位作者 David Olakanmi Olaniyi Francis Adole Agada Bamidele Musbau Adeniran Latifat Ronke Owoade 《World Journal of Nuclear Science and Technology》 CAS 2024年第2期118-130,共13页
Timer error as well as its convention is very important for dose accuracy during irradiation. This paper determines the timer error of irradiators at Secondary Standard Dosimetry Laboratory (SSDL) in Nigeria. The irra... Timer error as well as its convention is very important for dose accuracy during irradiation. This paper determines the timer error of irradiators at Secondary Standard Dosimetry Laboratory (SSDL) in Nigeria. The irradiators are Cs-137 OB6 irradiator and X-ray irradiators at the Protection level SSDL;and Co-60 irradiator at the Therapy Level SSDL. PTW UNIDOS electrometer and LS01 Ionization chamber were used at the Protection Level to obtain doses for both Cs-137 OB6 and X-ray irradiators while an IBA farmer type ionization chamber and an IBA DOSE 1 electrometer were used at the Protection Level SSDL. Single/multiple exposure method and graphical method were used in the determination of the timer error for the three irradiators. The timer error obtained for Cs-137 OB6 irradiator was 0.48 ± 0.01 s, the timer error for the X-ray irradiator was 0.09 ± 0.01 s while the timer error obtained for GammaBeam X200 was 1.21 ± 0.04 s. It was observed that the timer error is not affected by source to detector distance. It was also observed that the timer error of Co-60 Gamma X200 irradiator is increasing with the age of the machine. Source to detector distance and field size do not contribute towards the timer error of the irradiators. The timer error of the Co-60 Gamma X200 irradiator (the only irradiator among the irradiators with a pneumatic system) increases with the age of the irradiator. 展开更多
关键词 Timer Error Irradiation SSDL Irradiators dose accuracy
下载PDF
上一页 1 下一页 到第
使用帮助 返回顶部