The measurement errors due to nonparallelness of specimen faces in the ultrasonic velocity and attenuation coefficient measured by the pulse reflection method in VHF range have been investigated theoretically, and the...The measurement errors due to nonparallelness of specimen faces in the ultrasonic velocity and attenuation coefficient measured by the pulse reflection method in VHF range have been investigated theoretically, and the theoretical expressions for estimating these measurement errors are given. It has also been shown that, the attenuation coefficient error is depending on both the nonparallelness angle of specimen faces and the ultrasonic frequency, but the velocity error is only depending on the former. Furthermore, for pure silica glass specimen it is estimated that, in VHF range in order to insure that the attenuation coefficient and velocity errors due to nonparallelness of specimen faces are less than 10% and 0.01%, respectively, the nonparallelness angle of specimen faces must be less than 10 s and 40 s correspondingly.展开更多
文摘The measurement errors due to nonparallelness of specimen faces in the ultrasonic velocity and attenuation coefficient measured by the pulse reflection method in VHF range have been investigated theoretically, and the theoretical expressions for estimating these measurement errors are given. It has also been shown that, the attenuation coefficient error is depending on both the nonparallelness angle of specimen faces and the ultrasonic frequency, but the velocity error is only depending on the former. Furthermore, for pure silica glass specimen it is estimated that, in VHF range in order to insure that the attenuation coefficient and velocity errors due to nonparallelness of specimen faces are less than 10% and 0.01%, respectively, the nonparallelness angle of specimen faces must be less than 10 s and 40 s correspondingly.