A prototype of large-area position sensitive neutron detector was designed and constructed according to the requirements of the Small-Angle Scattering spectrometer of China Spallation Neutron Source(CSNS).The detector...A prototype of large-area position sensitive neutron detector was designed and constructed according to the requirements of the Small-Angle Scattering spectrometer of China Spallation Neutron Source(CSNS).The detector was based on the 3He neutron convertor and MWPC with an effective area of 650 mm×650 mm.A prototype was completed and tested with 55Fe X-ray.The high-pressure vessel was designed and constructed with high-strength aluminum alloy.A position resolution of about 4.6mm×2.3 mm(FWHM)and efficiency>65%for neutrons with wavelength of 1.8?was determined after the operational gas filled.展开更多
基金supported by the National Natural Science Foundation of China (Grant No.11127508)
文摘A prototype of large-area position sensitive neutron detector was designed and constructed according to the requirements of the Small-Angle Scattering spectrometer of China Spallation Neutron Source(CSNS).The detector was based on the 3He neutron convertor and MWPC with an effective area of 650 mm×650 mm.A prototype was completed and tested with 55Fe X-ray.The high-pressure vessel was designed and constructed with high-strength aluminum alloy.A position resolution of about 4.6mm×2.3 mm(FWHM)and efficiency>65%for neutrons with wavelength of 1.8?was determined after the operational gas filled.