BACKGROUND During the perianesthesia period,emergency situations threatening the life and safety of patients can occur at any time.When dealing with some emergencies,occasional confusion is inevitable.CASE SUMMARY Thi...BACKGROUND During the perianesthesia period,emergency situations threatening the life and safety of patients can occur at any time.When dealing with some emergencies,occasional confusion is inevitable.CASE SUMMARY This case report describes the rare situation wherein a surgeon inadvertently detached the inflatable tube of an endotracheal tube during a tonsillectomy,and positive pressure ventilation could not be provided.While reintubation may increase the risk of respiratory tract infection and aspiration,patients with a difficult airway might die due to apnea.The best treatment method is to optimize the damaged tracheal tube junction to avoid secondary intubation and ensure patient safety.An intravenous needle and cannula were used to repair the damaged gap in the current case.Following the repair,the anesthesia machine showed no indication of low tidal volume,and there was no deflation of the endotracheal tube cuff.Subsequently,the patient was transferred to the postanesthesia recovery room,and the tracheal tube was removed with satisfactory results.CONCLUSION Using an intravenous needle to repair a break in the inflatable tube surrounding an endotracheal tube is safe and reliable.展开更多
文摘BACKGROUND During the perianesthesia period,emergency situations threatening the life and safety of patients can occur at any time.When dealing with some emergencies,occasional confusion is inevitable.CASE SUMMARY This case report describes the rare situation wherein a surgeon inadvertently detached the inflatable tube of an endotracheal tube during a tonsillectomy,and positive pressure ventilation could not be provided.While reintubation may increase the risk of respiratory tract infection and aspiration,patients with a difficult airway might die due to apnea.The best treatment method is to optimize the damaged tracheal tube junction to avoid secondary intubation and ensure patient safety.An intravenous needle and cannula were used to repair the damaged gap in the current case.Following the repair,the anesthesia machine showed no indication of low tidal volume,and there was no deflation of the endotracheal tube cuff.Subsequently,the patient was transferred to the postanesthesia recovery room,and the tracheal tube was removed with satisfactory results.CONCLUSION Using an intravenous needle to repair a break in the inflatable tube surrounding an endotracheal tube is safe and reliable.